Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: DAD and MWD Detectors

In addition to (or different from) the standard detector commands (see Commands for Controlling Dionex Devices  Dionex Detectors) and the general commands for the 1100/1200 HPLC system (see  Agilent 1100/1200 HPLC System: General), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):

UV_VIS Channel Commands

Property

Description

AnalogAttenuation

Attenuation of the analog output (signal value that corresponds to the maximum output voltage). Values: -10 (º 2^-10 AU) - 1(º 2 AU); default: 0

Note: logarithmic scale

AnalogOffset

Zero offset of the analog output in % of the full output range. Values: 1 - 99%; default: 5

AnalogRange

Voltage range of the analog output (may be set to 0.1 or 1.0V only).

Bandwidth

Signal bandwidth. Values: 2 - 400 nm

MaxAutoStep

Maximum signal and spectra data rate if Step = Auto. Values: 0.1 s - 5.1 s; default: 0.1x

RefBandwidth

Reference bandwidth (set to 0 if RefWavelength = 0). Values: 0; 2 - 400 nm

RefWavelength

Reference wavelength (0 means: not used). Values: 0; 190 - 950 nm

Step

Signal and spectra data rate (If you set the step, you must set the PeakWidth too, see below). Values: 0.05 - 3.2 s; default: 0.4 s

Wavelength

Signal wavelength. Values: 190 - 950 nm

 

The relation between the expected PeakWidth and the corresponding Step is as follows:

Expected peak width [min]

0.00

0.01

0.02

0.05

0.10

0.20

0.40

0.85

Corresponding step [s]

0.05

0.05

0.10

0.20

0.40

0.80

1.60

3.20

 

3DFIELD Channel Commands (DAD only)

Property

Description

BunchWidth

Bandwidth of the 3D wavelength. Values: 0.01 - 100 nm; default: 2 nm

MaxWavelength

Maximum 3D wavelength. Values: 190 - 950 nm; default: 400

MinWavelength

Minimum 3D wavelength. Values: 190 - 950 nm; default: 190

Other Special Detector Commands

Property

Description

AutoActivateUV_Lamp

If set to On, the deuterium lamp is automatically turned on upon power-up. Values: On, Off (default)

AutoactivateVisible_
Lamp

If set to On, the Tungsten lamp is automatically turned on right after power-up of the detector. Values: On, Off (default)

Autozero

 

Performs Autozero. The command is performed immediately, when issued manually, or at a Program time <0.000 min. With a program time of 0.000, the command is performed before the injection program; with a program >0.000, the command is performed at this program time.

DemoFileName

File name of the demo file. (May contain %d that will be replaced by the sample number)

PeakWidth

 

Expected minimum peak width in [min] (If you set the PeakWidth, you must also set the Step: see above).

Values (Agilent 1100/1200 DAD): 0.00, 0.01, 0.02, 0.05, 0.10, 0.20, 0.20, 0.40, 0.85.

Values (Agilent 1200 DAD SL): 0.00, 0.0025, 0.0050, 0.01, 0.02, 0.05, 0.10, 0.20, 0.20, 0.40, 0.85.

SlitWidth

Width of the micro slit (also MWD). A small slit allows using the fine structures of the UV spectrum. A wide slit provides higher sensitivity. Values: 1 - 16 nm; default: 4 nm

UV_Lamp

Reports the state of the deuterium lamp. Values: On, Off (default)

UV_LampRequired

If set to On: The detector is not ready for operation if the deuterium lamp is off. Values: On, Off (default)

Visible_Lamp

Tungsten lamp state. Values: On, Off (default)

Visible_LampRequired

If set to On: The detector is not ready for operation if the Tungsten lamp is off. Values: On, Off (default)

Counters

The values are read from the module on connect.

Property

Description

FilterMovements

Counter for the movements of the holmium oxide filter.

Set the UpperLimit to 0 to disable the associated limit test.

SlitMovements

Slit movement counter.

Set the UpperLimit to 0 to disable the associated limit test.

UvLampIgnitions

UV lamp ignition counter.

Set the UpperLimit to 0 to disable the associated limit test.

UvLampAge

Total burn time of the UV lamp [h]

Set the UpperLimit to 0 to disable the associated limit test.

UvLampOntime

Burn time of the UV lamp since the last time it was turned on.

Set the UpperLimit to 0 to disable the associated limit test.

VisLampIgnitions

VIS lamp ignition counter.

Set the UpperLimit to 0 to disable the associated limit test.

VisLampAge

Total burn time of the VIS lamp [h]]

Set the UpperLimit to 0 to disable the associated limit test.

VisLampOntime

Burn time of the VIS lamp since the last time it was turned on.

Set the UpperLimit to 0 to disable the associated limit test.

UV Calibration Commands

Property

Description

Adjust_Calibration

Adjusts the wavelength calibration using the results of a previous wavelength calibration (absolute deviation).

D2_Alpha_Deviation_
Absolute
or

D2_Beta_Deviation_
Absolute

Current deviation of D2 alpha or D2 beta line from factory calibration.

Valid only after the Test_Calibration command has been completed.

D2_Alpha_Deviation_
Correction
or

D2_Beta_Deviation_
Correction

Current calibration for the D2 alpha or D2 beta line. Values: -2.000 - 2.000 nm (read from module on connect)

D2_Alpha_Deviation_
Current
or

D2_Beta_Deviation_
Current

Current deviation of the D2 alpha or D2 beta line, considering current wavelength calibration. Values: -2.000 - 2.000 nm Valid only after the Test_Calibration command has been completed.

Test_Calibration

Performs a wavelength calibration test as described in the Agilent VWD Detector Reference Manual.

You need to remove the flow cell before executing this command. In addition, the UV lamp must be on.

Diagnostic Commands and Properties

 

Command

Description

CellTest

Performs a cell test.

CellTestEvaluation

Evaluates the cell test by averaging the spectra of all test data.

CellTestInfo

Indicates whether the cell test in the next step is performed with or without cell.

DarkCurrentTest

Performs a dark current test.

DarkCurrentTestEvaluation

Evaluates the dark current test by averaging the minimum and maximum values of all selected spectra ranges.

FilterTest

Performs a holmium filter test.

FilterTestEvaluation

Evaluates the holmium filter test.

IntensityTest

Performs an intensity test.

IntensityTestEvaluation

Evaluates the intensity test by calculating the minimum and maximum values of the 1024 diodes for all test data.

 

 

For more information about the diagnostic functions for the DAD detector, refer to:

 Agilent 1100 HPLC System Diagnostics: Dark Current Test

 Agilent 1100 HPLC System Diagnostics: Cell Test

 Agilent 1100 HPLC System Diagnostics: Filter Test

 Agilent 1100 HPLC System Diagnostics: Intensity Test

 