Bandwidth

 

Instrument Type:

Detector

Type:

Integer

Value Range:

Detector-dependent

Default:

0 [nm]

Related Commands:

Wavelength

RefBandwidth (Reference Bandwidth)

BunchWidth

Description:

The Bandwidth command specifies the optical bandwidth [in nm] at which a chromatogram (UV VIS channel) is recorded. In general, this corresponds to the Optical Resolution of a detector.

Function:

You can simulate (i.e., increase) the bandwidth by averaging several single photodiode signals. This process is known as Photodiode Bunching. Averaging is performed symmetrically to the selected wavelength. Thus, at a bandwidth of 31 nm and a wavelength of 255 nm, the signals of all photodiodes between 240 and 270 nm are averaged.

 Tip:

Changing the bandwidth can often increase the sensitivity. Quadrupling the bandwidth almost halves the noise. However, in this case, linearity usually decreases.