Commands and Tips for Thermo Scientific
Devices
Thermo Scientific TriPlus Sampler: General Commands
The following general commands and properties are available for the TriPlus autosampler. Please note that the display Filter level determines which commands and properties are displayed.
Also, refer to the TriPlus Operating Manual supplied by Thermo Scientific for more information on controlling the instrument.
Note:
Command and property names used by Thermo Scientific may slightly differ from Chromeleon terms. For details please refer to column "Thermo Scientific" in the tables below.
Configuration and Status Commands/Properties:
Command/Property |
Range/Values |
Description |
FirmwareVersion |
- |
Indicates the sampler's firmware version (read-only). |
HardwareVersion |
- |
Indicates the sampler's hardware version (read-only). |
NeedleLength |
50-80 mm |
Indicates the length of the needle in [mm] (read-only). |
OperationMode |
Main, Virtual |
Displays the setting of the Virtual (Secondary) Device option configured on the General page in the Server Configuration (read-only). |
Ready |
Ready, NotReady |
Reports whether the sampler is ready for the next injection (read-only). Also refer to property Status. Note: To be ready for injection, the Status property of the related temperature modules must be 'Ready', too. |
SamplerType |
Unknown, GC liquid, Headspace, HPLC, SPME (Solid Phase Microextraction) |
Indicates the configured sampler type (read-only). |
Status |
|
Indicates the sampler's status (read-only). |
Syringe |
0-5000 µl |
Indicates the capacity of the syringe in [µl]. |
TrayTypePrimary / TrayTypeSecondary |
|
Reports the type of the primary or secondary sample tray (read-only). |
|
None Tray_1_54 Tray_55_108 Tray_1_150 Tray_151_300 Tray_1_192 Tray_193_384 Tray_1_768 Tray_769_2304 Tray_1_33 Tray_34_66 Tray_1_96 Tray_97_192 |
Unknown, not defined or deactivated Tray holding 54 vials (1-54) Tray holding 54 vials on second tray holder (55-108) Tray holding 150 vials (1-150) Tray holding 150 vials on second tray holder (151-300) Tray for 2x96 well plates (1-192) Tray for 2x96 well plates on second tray holder (193-384) Tray for 2x384 well plates (1-768) Tray for 2x384 well plates on second tray holder (769-2304) Cooled tray holding 33 vials (1-33) Cooled tray holding 33 vials on second tray holder (34-66) Cooled tray holding 96 vials (1-96) Cooled tray holding 96 vials on second tray holder (97-192) |
General Commands and Properties:
Command/Property |
Thermo Scientific |
Range/Values |
Description |
GCStartMode |
GC Synchro Start |
TriPlus AS version: Standard Anticipated Delayed None
TriPlus HS version: Normal Anticipated |
Specifies the start mode of the GC. The start mode determines when the sampler sends the start signal to the GC. For details, refer to the TriPlus Operating Manual. |
Inject |
|
Position Volume Blank |
Injects the sample. If you create a user-defined program, e.g., in the Program Wizard, use the Inject command to start the program. Notes: The Position and Volume values can only be used for manual injections. Only one injection per program is allowed. Injection is allowed only at t=0.000 min. |
InjectMode |
Injection Mode (TriPlus AS)
Incubation Mode (TriPlus HS) |
TriPlus AS version:
TriPlus HS version:
|
Specifies the injection mode. Refer to the TriPlus Operating Manual for detailed information on injection modes. |
InjectionsPerVial |
Repetition |
1-10 |
Specifies the number of MHE injections to be performed from the same vial. Set this parameter to 1, if InjectMode is 'Single' or 'Progressive'. Otherwise, an error message is displayed. Note: This parameter is read-only for TriPlus AS samplers. |
Injector |
|
InjectorA/B/C/D |
Specifies which injector (GC inlet) is used for injecting the sample. Notes: If an injector is selected which has not been installed, an error message is displayed. If a new injector is set up, the corresponding injector type is displayed under InjectorType. |
InjectorType |
|
|
Displays the type of the associated injector which has been specified under Injector (read-only). |
Position |
|
1-2304 |
Indicates the vial position. |
Reset |
|
|
Aborts a running sequence and sets the sampler's X, Y, Z axes and plunger to the initial condition. |
Volume |
Sampler volume (TriPlus AS)
Sample draw (TriPlus HS) |
0.1-15 x 250 µl (TriPlus AS)
0-4.5 ml x 16 ml (TriPlus HS) |
Specifies the injection volume. The maximum injection volume may not exceed 90% of the syringe capacity. Note: The injection volume may differ from the volume per load, if InjectMode is 'Enrichment' . Also refer to Commands for Enrichment Injections. |
Commands for Controlling the Sampler's "Rapid Mode" and "Home After Sample" Functions:
Command/Property |
Thermo Scientific |
Range/Values |
Description |
RapidMode |
Rapid Mode Advanced Parameters |
WaitAfterRinse WaitAfterBubble WaitAtInjectPort WaitAfterRinse_HomeAfterSample WaitAfterBubble_HomeAfterSample WaitAtInjectPort_HomeAfterSample |
Specifies the rapid mode, if RapidModeCtrl is enabled. Notes: Rapid mode settings can only be specified on the control panel or in the Commands dialog box, but not in the PGM file. To specify settings, make sure that the sampler is idle and no batch is running. |
RapidModeCtrl |
Rapid mode |
ON/OFF |
Enables or disables the Rapid Mode function for GC liquid modes. Notes: Rapid mode settings can only be specified on the control panel or in the Commands dialog box, but not in the PGM file. To specify settings, make sure that the sampler is idle and no batch is running. |
Common Wash Commands:
Command/Property |
Thermo Scientific |
Range/Values |
Description |
PostWash |
Post Injection Cycles |
0-15 |
Specifies the number of wash cycles after injection. |
PostWashSolvent1/2/3/4 |
Solvents |
|
Specifies the solvent(s) to be used during one post-injection wash cycle (see also parameter PreWashSolventSequence.) For more information also refer to the TriPlus Operating Manual. |
PostWashSolventSequence |
|
|
Reports the order of solvents (A, B, C, D) used within a postwash cycle, determined by the parameters WashMode and PostWashSolvent1/2/3/4 (read only). A solvent sequence may look like this: 'ABD-', whereas '-' means that no solvent is selected for this step. To obtain this solvent sequence you have to select: WashMode = Multiple Solvents PreWashSolvent1 = A PreWashSolvent2 = B PreWashSolvent3 = D PreWashSolvent4 = None |
PostWashVolume |
Pre-injection solvent volume |
0-250 µl (TriPlus AS)
0-0.9*syringe volume (TriPlus HS)
|
Specifies the solvent volume to be drawn for the wash cycle after injection. |
Wash |
|
|
Starts a manual wash procedure. During washing, the settings of the post-injection wash parameters are taken into account (also refer to parameters PostWash, PostWashSolvent1/2/3/4, PostWashSolventSequence, and PostWashVolume). |
WashMode |
Solvents |
SingleSolvent, MultipleSolvents |
Specifies the number of solvents used during the wash procedure. Also refer to parameters PreWashSolventSequence and PostWashSolventSequence. |
Commands for Enrichment Injections:
Command/Property |
Thermo Scientific |
Range/Values |
Description |
InjectionVolume |
Sample volume (TriPlus AS)
Sample draw (TriPlus HS) |
0-250 µl (TriPlus AS)
0-4.5 ml (TriPlus HS)
|
Reports the actual injection volume per sample load given by Volume/LoadsPerInject (read-only). Note: The InjectionVolume parameter is used to check the sample volume against the capacity of the installed syringe. |
LoadsPerInject |
Enrichment injection repetitions |
1-15 (TriPlus AS)
1-16 (TriPlus HS) |
Sets the number of sample enrichments during an injection cycle. Note: The sample volume per enrichment is Volume/LoadsPerInject. |
LoadRepetitionTime |
Time between enrich.Inj. (TriPlus AS)
Enrichment delay (TriPlus HS) |
1-600s (TriPlus AS)
0.1-60.0 min (TriPlus HS) |
Specifies the delay between one load/inject cycle and the next during enrichment mode. |
Commands for Components with Temperature Control:
Depending on the installed instrument version, the TriPlus sampler is equipped with a variety of thermostatted components such as temperature controlled primary tray, temperature controlled secondary tray, headspace syringe needle, and incubation oven (agitator). To control these components the following commands are provided:
Command/Property |
Range/Values |
Description |
Ctrl |
On, Off |
Turns the temperature control On or Off during a batch run. |
LowerLimit |
Depends on the component |
Sets the minimum temperature for Nominal and StandbyNominal parameters. |
Nominal |
Depends on the component and UpperLimit and LowerLimit parameter settings. |
Sets the target value of the temperature control during a batch run and activates the Ctrl control parameter (Ctrl = On). |
StandbyCtrl |
On, Off |
Turns the temperature control for standby mode On or Off. |
StandbyNominal |
Depends on the component and UpperLimit and LowerLimit parameter settings. |
Sets the target value of the temperature control for standby mode and activates the StandbyCtrl control parameter (StandbyCtrl = On). |
Status |
Ready, Unknown, AboveNominal, BelowNominal, SafetyCutOff |
Indicates the component's status (read-only). Note: If the component's status is not Ready, the sampler's status is changed from Ready to NotReady. |
UpperLimit |
Depends on the component |
Sets the maximum temperature for Nominal and StandbyNominal parameters. |
Value |
Depends on the component |
Reports the current temperature of the component (read-only). |
Note:
To control the temperature of a thermostatted component, the device must be connected to the sampler which is configured as "Main" device. If connected to the "Virtual" sampler, the component's temperature can only be monitored.
For information on the specific commands and properties for the TriPlus AS and HS sampler versions, refer to:
Thermo Scientific TriPlus Sampler: Commands for the TriPlus AS
Thermo Scientific TriPlus Sampler: Commands for the TriPlus HS
For program examples to control the TriPlus samplers, refer to:
Creating a Program for Liquid Injections
Creating a Program for Headspace Analysis
The Administrator Help section provides installation details; refer to Hardware Installation:
For information about how to install the sampler, refer to Thermo Scientific TriPlus Sampler: Installation.
If you want to operate the sampler with two GCs, refer to Thermo Scientific TriPlus Sampler: Dual Operation.