Thermo
Scientific Detectors
Thermo Scientific Vanquish Diode Array Detector
The commands listed below are available for the Vanquish DAD detector. Please note that the display Filter level and detector type determine which commands and properties are displayed.
3DFIELD channel
Command/Property |
Description |
Use this command to start/stop data acquisition. |
|
Use this property to specify the range that is averaged when collecting 3D data (similar to the Bandwidth of a UV channel). A higher BunchWidth reduces the required data storage and improves the signal-noise-ratio, but reduces the Optical Resolution. Note: Only integer values are accepted. If you enter a decimal value, Chromeleon automatically rounds to an integer value. |
|
MaxWavelength |
Use this property to set the maximum wavelength for recording spectra. Narrows the range in which the 3D field is recorded (also see MinWavelength). |
MinWavelength |
Use this property to set the minimum wavelength for recording spectra. Narrows the range in which the 3D field is recorded (also see MaxWavelength). |
Use this property to average several photodiode signals of the reference signal. It is not necessary to select a reference bandwidth unless a reference wavelength is set. |
|
Use this property to correct the 3D field absorption values. The selected reference wavelength should be in an area of the spectrum where little absorption occurs. |
|
Indicates the retention time of the signal. |
UV channel
Command/Property |
Description |
Use this command to start/stop data acquisition. |
|
Use this property to specify the optical bandwidth at which a chromatogram is recorded. |
|
Indicates the signal slope, that is, the difference between the current value and the value one second ago. |
|
Use this property to average several photodiode signals of the reference signal. It is not necessary to select a reference bandwidth unless you enter a RefWavelength. |
|
Use this property to correct the absorption values of the wavelength that has been selected for the analysis. The selected reference wavelength should be in a quiet area of the spectrum where little absorption occurs. |
|
Indicates the signal's retention time . |
|
Signal |
Indicates the current signal value [mAU] . |
Use this property to set the wavelength for recording the signal on the UV channel to the wavelength with the absorbance maxima for the analytes of interest. |
Equilibration commands and properties
Use the equilibration commands and properties only for the SmartStartup feature.
Command/Property |
Description |
|
Drift |
Provides information about the detector Drift, determined by equilibration. |
|
*.UpperLimit |
Use this property to set the upper drift limit [mAU/h]. Set to 0 to turn drift equilibration off. |
|
*.Value |
Indicates the current drift value in mAU/h. The value is calculated as soon as at least a one-minute segment was measured. It is updated with each new recorded segment and averaged over the last 5 minutes. |
|
DriftEquilibration |
Indicates the status/result of the drift measurement. |
|
DriftStatus |
Indicates the status/result of the drift measurement. |
|
|
N/A |
Equilibration has not started. |
|
Measuring |
Equilibration is running, but the amount of data is not yet sufficient to report a result. |
|
NotReady |
Equilibration is running or has finished. The values are within the specified limits. |
|
Good |
Equilibration is running. The current data is outside the specified limits. |
|
Failed |
Equilibration is completed. The result is outside the specified limits. |
|
NotTested |
The limit was set to 0, drift measurement is disabled. |
Equilibration |
Provides information about the total status of the equilibration: OK or NotOK. |
|
Noise |
Provides information about the detector Noise. |
|
*.UpperLimit |
Use this property to set the upper noise limit in [mAU]. Set to 0 to turn noise equilibration off. |
|
*.Value |
Indicates the current noise. The value is calculated as soon as at least a one-minute segment was measured. It is updated with each new recorded segment and averaged over the last 5 minutes. |
|
NoiseEquilibration |
Indicates the status/result of the noise measurement: OK or NotOK. |
|
NoiseStatus |
Indicates the status/result of the noise measurement. For the available values, see DriftStatus. |
|
StartEquilibration |
Use this command start equilibration (noise/drift measurement). |
General detector properties (read-only)
Command/Property |
Description |
Alarm |
Indicates whether the detector reports an alarm. The detector will not clear an alarm automatically, manual intervention is required. Call MuteAlarm to mute the audible alarm. |
AmbientTemp |
Indicates the temperature [°C] outside the optical system. |
ChannelIntensity1 |
Indicates an intensity value. |
ChannelIntensity2 |
Indicates an intensity value. |
ChannelIntensity3 |
Indicates an intensity value. |
ChannelIntensity4 |
Indicates an intensity value. |
ChannelIntensity5 |
Indicates an intensity value. |
ChannelIntensity6 |
Indicates an intensity value. |
ChannelIntensity7 |
Indicates an intensity value. |
ChannelIntensity8 |
Indicates an intensity value. |
ChannelIntensity9 |
Indicates an intensity value. |
ChannelIntensity10 |
Indicates an intensity value. |
DriverId |
Indicates the unique driver identifier. |
DriverManufacturer |
Indicates the manufacturer of the driver. |
DriverVersion |
Indicates the version of the driver. |
ExceptionLogClear |
Use this command to clear the error log of the device. Note: An error log is created if the firmware has crashed during an operation. The firmware sends all error log entries to the driver at connect time to be reported in the daily audit trail. You get a warning message about the error and are asked to send the affected audit trail to hardware service for inspection if this happens more often, and to use this command to clear the internal logs. |
FirmwareVersion |
Indicates the firmware version of the module. |
FlowCellOperationTime |
Indicates the time that the flow cell has been operated. |
IlluminatedVolume_Flowcell |
Indicates the illuminated volume of the flow cell. |
LamphouseTemp |
Indicates the lamp house temperature [°C]. |
LastUVLampChangeDate |
Indicates when the UV lamp was replaced last. |
Leak |
Indicates whether the leak sensor has reported a leak. |
LeftDoor |
Indicates whether the compartment door is Open or Closed. |
MaxPressure_Flowcell |
Indicates the pressure limit of the flow cell. |
ModelNo |
Indicates the model number. |
NotReadyCauses |
Indicates why the module is not yet ready for the next analysis. |
OperatingHours |
Indicates the operating hours since the detector was operated for the first time. |
OpticalPath_Flowcell |
Indicates the optical path length of the flow cell. |
PowerOnTime |
Indicates the time since the detector was powered-up last. |
PreviousUVLampOperation |
Indicates the operating hours of the previously installed UV lamp. |
ProductNumber_Flowcell |
Indicates the product number of the flow cell. |
Qualification.LastDate |
Indicates the date when detector qualification was performed last. |
Qualification.LastOperator |
Indicates the name of the person who performed the detector qualification. |
Ready |
Indicates that the module is ready for the next injection. |
RightDoor |
Indicates whether the compartment door is Open or Closed. |
SerialNo |
Indicates the serial number of the device. |
SerialNo_Flowcell |
Indicates the serial number of the flow cell. |
SlitChangeCount |
Indicates the number of times the slit has been changed. |
SpectrographTemp |
Indicates the temperature of the spectrograph. |
Type_Flowcell |
Indicates the flow cell type. |
UvLamp_ProductNumber |
Indicates the product number of the UV lamp. |
UvLamp_SerialNo |
Indicates the serial number of the UV lamp. |
UVLampIgnitions |
Counts the total number of successful UV lamp ignitions. |
UVLampIntensity |
Indicates the UV lamp intensity at 254 nm in counts per second. |
UVLampOnTime |
Counts the operating hours of the UV lamp since the lamp was ignited last. |
UVLampOperationTime.Value |
Indicates the total operating hours of the UV lamp. |
WavelengthPeak1Measured |
Indicates the measured peak of the first wavelength. |
WavelengthPeak1Theoretical |
Indicates the theoretical peak of the first wavelength. |
WavelengthPeak2Measured |
Indicates the measured peak of the second wavelength. |
WavelengthPeak2Theoretical |
Indicates the theoretical peak of the second wavelength. |
WavelengthPeak3Measured |
Indicates the measured peak of the third wavelength. |
WavelengthPeak3Theoretical |
Indicates the theoretical peak of the third wavelength. |
WavelengthPeak4Measured |
Indicates the measured peak of the fourth wavelength. |
WavelengthPeak4Theoretical |
Indicates the theoretical peak of the fourth wavelength. |
WavelengthPeak5Measured |
Indicates the measured peak of the fifth wavelength. |
WavelengthPeak5Theoretical |
Indicates the theoretical peak of the fifth wavelength. |
WavelengthPeak6Measured |
Indicates the measured peak of the sixth wavelength. |
WavelengthPeak6Theoretical |
Indicates the theoretical peak of the sixth wavelength. |
WavelengthPeak7Measured |
Indicates the measured peak of the seventh wavelength. |
WavelengthPeak7Theoretical |
Indicates the theoretical peak of the seventh wavelength. |
General detector commands
Command |
Description |
AutoactivateUV_Lamp |
Use this property to set the UV lamp. When set to On, the UV lamp is turned on automatically whenever the detector is powered up. |
Use this command to perform automatic null balancing (= the current detector signal is interpreted as 0). |
|
BuzzerVolume |
Use this property to specify the volume of the buzzer. |
CalibrationReset |
Use this command to reset all calibration settings to factory defaults. |
Connect |
Use this property to connect the device to the data system, enabling computer control. |
Connected |
Indicates whether the device is connected to the data system that is under computer control. |
Data_Collection_Rate |
Enter the rate [Hz] at which Chromeleon collects digital data from the detector. |
DemoFileName |
Use this property to enter the file name of the demo file (may contain %d, which will be replaced by the sample number). |
Disconnect |
Use this command to disconnect the device from the data system and enables manual control from the keypad of the device. |
LampCover |
Indicates whether the lamp cover is attached or not. |
LeakSensorMode |
Use this property to activate or deactivate leak detection.
|
MuteAlarm |
Use this property to mutes the audible alarm. The command has no effect if the alarm state is Off. |
PeakWidth |
Use this property to set the narrowest peak width at half height. |
QualificationDone |
Use this command when a qualification was done to update the associated information. |
Parameter: At |
desired wavelength |
Parameter: doubleValue |
double rounded as int (optional) |
Parameter: index |
Index (optional) |
Parameter: intValue |
intValue (optional) |
Parameter: String |
String |
Parameter: strValue |
strValue (optional) |
Parameter: token |
token (optional) |
Parameter: Which |
Defines which Thermo Scientific sensor has been added. |
Reset |
Use this command to reset the device to its initial condition, as attained after power-up. |
ResponseTime |
Use this property to specify the Response Time for the signal. The response time is the time required for the detector signal to rise from 10% to 90% in response to a signal increase. A longer response time reduces the noise. However, if the selected response time is too long, this can result in reduced peak heights and asymmetrical peak shapes. The value calculated by Chromeleon is only a recommended value. You are free to select a different value as appropriate for your application. Tip: The response time multiplied by the data collection rate must not be smaller than 2 or greater than 20. |
ServiceDone |
Use this command when a service was done to update the associated information. |
SetFactoryDefaults |
Use this property to restore the Thermo Scientific standard settings. |
Shutter |
Use this property to specify whether the shutter blocks the light path. |
SlitWidth |
Use this property to specify the slit width. A wide slit allows more light to fall on the array. A smaller slit results in a higher optical resolution. |
UV_Lamp |
Use this property to indicate the state of the UV lamp. To turn on the deuterium lamp, set UV_Lamp to On. |
UV calibration and validation commands
Command |
Description |
WavelengthCalibration |
Use this command to perform wavelength calibration via the deuterium lamp. Note: To ensure that this command is executed correctly, a Wait UV.Ready command is required after this command in the PGM File. |
WavelengthValidation
|
Use this command to perform an internal wavelength validation procedure using the holmium filter. Note: To ensure that this command is executed correctly, a Wait UV.Ready command is required after this command in the PGM File. |
System Wellness and Predictive Performance commands and properties
Command/Property |
Description |
LastUVLampChangeDate |
Indicates when the UV lamp was replaced last . |
Qualification |
Provides information about system qualification. |
*.Interval
|
Use this property to specify the interval (in days) when the next system qualification is due. |
*.WarningPeriod |
Use this property to specify the period during which a warning appears before the next qualification is due. |
*.GracePeriod |
Use this property to specify the period during which the detector may be operated after the due date for re-qualification. |
QualificationDone |
Use this command after qualification to update the related system data. |
Service |
Provides detector service information. |
*.LastDate |
Indicates the date of the last service . |
*.Interval |
Use this property to specify the interval (in days) after which a service is due. |
*.WarningPeriod |
Use this property to specify the period during which a warning appears before the next service is due. |
ServiceDone |
Use this command after the device has been serviced, select this command to update the related system information. |
SlitChangeCount |
Indicates the number of times that the slit width has been changed. |
UVLampOperationTime |
Provides information on the UV lamp's operating time. |
*.Limit |
Use this property to set the limit for the operating time of the UV lamp. An error message appears when the specified value is reached. |
*.Value |
Indicates the total operating time of the UV lamp in [h]. |
*.Warning |
Use this property to specify a time limit for issuing a warning with regard to the operation time. A warning appears when the specified value is reached. |
The following commands and properties are available only if the DAC plug-in module is installed and the DAC Board check box is selected in the Properties dialog for the detector (Detector tab page) in the Server Configuration program.
Property |
Description |
Analog1_Offset |
Use this property to adjust the zero position of the analog output when it is plotted. The value entered is a percentage of the full-scale analog output. The offset allows the evaluation device to plot the signal if it becomes negative. Note: The output voltage of the analog outputs cannot reach a minimum of exactly 0 V. Therefore, we recommend to set an offset of at least 5 %, even if only positive analog output signals are expected. This will enable you to also measure very small output signals down to 0 precisely. |
Analog1_Range |
Use this property to set the voltage range of the analog output signal. The range to use depends on the evaluation device to which the output is connected. |
Analog1_Resolution |
Use this property to set the resolution of the analog output signal. The resolution to use depends on the value expected for the application. |
Further Information
For an overview of the commands for all devices supported by Thermo Fisher Scientific, refer to Thermo Scientific: Commands and Tips.
For installation details, refer to Hardware Installation: Thermo Scientific Devices:
Vanquish - Diode Array Detector.