Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: VWD Detector

In addition to (or different from) the standard detector commands (see Commands for Controlling Dionex Devices  Dionex Detectors) and the general commands for the 1100 and 1200 HPLC systems (see  Agilent 1100/1200 HPLC System: General), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):

UV_VIS Channel Commands

Property

Description

AnalogAttenuation

Attenuation of the analog output (signal value that corresponds to the maximum output voltage). Values: -10 (º 2^-10 AU) - 1(º 2 AU); default: 0

Note: logarithmic scale

AnalogOffset

Zero offset of the analog output in % of the full output range. Values: 1 - 99%; default: 5

AnalogRange

Voltage range of the analog output (may be set to 0.1 or 1.0V only).

MaxAutoStep

Maximum signal and spectra data rate if Step = Auto. Values: 0.1 s - 5.1 s; default: 0.1x

PeakWidth

Expected minimum peak width in [min] (The setting affects the Step: see below).

Values (Agilent 1100/1200): 0.00, 0.01, 0.02, 0.05, 0.10, 0.20, 0.20, 0.40, 0.85 (read from module upon connect)

Values (Agilent 1200 VWD SL): 0.00, 0.00125, 0.0025, 0.005, 0.01, 0.02, 0.05, 0.10, 0.20, 0.20, 0.40, 0.85 (read from module upon connect)

Step

Signal and spectra data rate (The setting affects the peak width: see below). Values: 0.05 - 3.2 s; default: 0.4 s

Wavelength

Signal wavelength. Values: 190 - 950 nm

 

The relation between the expected peak width and the corresponding step is as follows:

Expected peak width [min]

0.000

0.005

0.010

0.025

0.050

0.100

0.200

0.400

Corresponding step [s]

0.073

0.073

0.073

0.073

0.146

0.291

0.582

1.165

Other Special Detector Commands

Property

Description

AutoActivateUV_Lamp

If set to On, the deuterium lamp is automatically turned on upon power-up. Values: On, Off (default)

AutoactivateVisible_
Lamp

If set to On, the Tungsten lamp is automatically turned on right after power-up of the detector. Values: On, Off (default)

Autozero

 

Performs Autozero. The command is performed immediately, when issued manually, or at a Program time <0.000 min. With a program time of 0.000, the command is performed before the injection program; with a program >0.000, the command is performed at this program time.

DemoFileName

File name of the demo file. (May contain %d that will be replaced by the sample number)

Scan

Records an absorbance spectrum. Use the ScanStartWvl and ScanEndWvl parameters to define the wavelength range to be scanned (190-600 nm). The ScanStepSize defines the step size for the scan (1 ... 10 nm).

ScanBaseline

Records an intensity spectrum of the baseline (blank spectrum). Use the ScanStartWvl and ScanEndWvl parameters to define the wavelength range to be scanned (190-600 nm). The ScanStepSize defines the step size for the scan (1 ... 10 nm). The baseline spectrum is saved in the detector's memory.

UV_Lamp

 

Reports the state of the deuterium lamp. Values: On, Off (default)

UV_LampRequired

 

If set to On: The detector is not ready for operation if the deuterium lamp is off. Values: On, Off (default)

Counters

The values are read from the module on connect.

Property

Description

ReferenceDiodeExposure

Counter of the reference diode exposure.

Set the UpperLimit to 0 to disable the associated limit test.

SampleDiodeExposure

Counter of the sample diode exposure.

Set the UpperLimit to 0 to disable the associated limit test.

UvLampIgnitions

UV lamp ignition counter.

Set the UpperLimit to 0 to disable the associated limit test.

UvLampAge

Total burn time of the UV lamp [h]

Set the UpperLimit to 0 to disable the associated limit test.

UvLampOntime

Burn time of the UV lamp since the last time it was turned on.

Set the UpperLimit to 0 to disable the associated limit test.

UV Calibration Commands

Property

Description

Adjust_Calibration

Adjusts the wavelength calibration using the absolute deviations determined by the test.

Test_Calibration

Performs a wavelength calibration test as described in the Agilent VWD Detector Reference Manual.

You need to remove the flow cell before executing this command. In addition, the UV lamp must be on.

Diagnostic Commands and Properties

 

Command

Description

CellTest

Performs a cell test.

CellTestEvaluation

Evaluates the cell test by averaging the spectra of all test data.

CellTestInfo

Indicates whether the cell test in the next step is performed with or without cell.

DarkCurrentTest

Performs a dark current test.

FilterTest

Performs a holmium filter test.

FilterTestEvaluation

Evaluates the holmium filter test.

IntensityTest

Performs an intensity test.

IntensityTestEvaluation

Evaluates the intensity test by calculating the minimum and maximum values of the entire wavelength range for all test data.

 

For more information about the diagnostic and scan functions for the VWD detector, refer to:

 Agilent 1100 HPLC System Diagnostics: Dark Current Test

 Agilent 1100 HPLC System Diagnostics: Cell Test

 Agilent 1100 HPLC System Diagnostics: Filter Test

 Agilent 1100 HPLC System Diagnostics: Intensity Test

Agilent 1100 HPLC System: Recording Absorbance Spectra (VWD Detector)

 

Note:

If the detector is equipped with a "B" firmware version (for example, B.06.14), it is not possible to run diagnostics tests in Chromeleon. Use the Agilent Lab Advisor instead.

 

The Administrator Help section provides installation instructions for the 1100 and 1200 HPLC systems; refer to Hardware Installation:

 Agilent 1100 HPLC System: Installation

 Agilent 1200 HPLC System: Installation