Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: Autosampler (G1329)

In addition to (or different from) the standard autosampler commands (see Commands for Controlling Dionex Devices  Dionex Autosamplers) and the general commands for the 1100 and 1200 HPLC systems (see  Agilent 1100/1200 HPLC System: General), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):

 

Property

Min

Max

Default

Description

DispSpeed

10 µl/min or
90 µl/min

1000 µl/min

200 µl/min

Speed for dispensing the sample

DrawSpeed

10 µl/min or
90 µl/min

1000 µl/min

200 µl/min

Speed for drawing the sample

Position

0

(Tray dependent)

n/a

Sample vial number

SampleHeight

-2.5 mm

35 mm

0.0 mm

Needle draw/eject position offset

Syringe

100 µl

900 µl

as configured

Syringe volume [µl]

Volume

0.0 µl

8000.0 µl

5.0 µl

Injection volume

WashVial

0

Greatest sample position (tray dependent)

0

Wash vial number
0: no wash

 

Command

Description

Dispense

Injects sample into the loop.

Draw

Draws sample from a vial into the syringe.

InitiateChangeGripper

Initiates the gripper change by moving the gripper to the front.

InitiateChangeNeedle

Initiates the needle change by removing the safety covering and moving the needle carrier to an easily accessible mounting position. In addition, the needle is raised to the uppermost position.

InitiateChangePiston

Initiates the piston change by releasing the spring on the proportioning unit.

Inject

Performs an injection (see note below).

Mix

Mixes the sample using the syringe.

Repeat/EndRepeat

Repeats the enclosed sample preparation steps.

TerminateChangeGripper

Terminates the gripper change by returning the gripper to the Home position.

TerminateChangeNeedle

Terminates the needle changes by returning the needle to the Home position.

TerminateChangePiston

Terminates the piston change by tensing the spring on the proportioning unit.

Valve

Switches the injection valve.

Wait

Waits a given time during the injection program.

 

Parameters of the commands

Command

Values

Description

Dispense,
parameter: Vial

SampleVial (current sample vial)
SampleVialPlus (current sample vial + n)
Vial (sample vial No. n)
Seat (needle seat)
Air

Position of the vial for dispense

Dispense/ Mix,
parameter: Speed

10 (or 90 µl) to 1000 µl/min
Maximum
Default (uses DispSpeed)

Speed for dispense or mix

Draw,
parameter: Source

SampleVial (current sample vial)
SampleVialPlus (current sample vial + n)
Vial (sample vial No. n)
Seat (needle seat)
Air

Position of the vial for draw

Draw,
parameter: Speed

10 (or 90) to 1000 µl/min
Maximum
Default (uses DrawSpeed)

Speed for draw

Draw/Dispense,
parameter: Offset

-2.5 to 35.0 mm
SampleHeight

Needle draw/eject position offset for draw or dispense

Draw/ Dispense/ Mix,
parameter: Volume

0.0 to 100.0 µl (or 900.0 µl)
Maximum
InjectVolume

Volume for draw, dispense, or mix

Needle,
parameter: Vial

SampleVial (current sample vial)
SampleVialPlus (current sample vial + n)
Vial (sample vial No. n)

Position to which the needle is moved

 

 Tips:

The Inject command sends a start impulse to all modules. Then the analysis will start with the following steps:

 

Three different injection programs are available:

Standard (injection volume from the respective position)

WashVial=None and no sample preparation commands

Standard (with needle wash)

WashVial=None and no sample preparation commands

User-defined injection program

If sample preparation commands are issued

 

Autosampler/Thermostat

Property

Min/Max

Default

Description

AutoActivateSetpoint

Off/On

On

If set to On, temperature control is automatically enabled right after power-up.

TempCtrl

Off/On

Off

Turns temperature control on or off. If set to On, temperature control is activated and the current temperature as shown under Temperature.Value is used as the temperature setpoint.

To define a different setpoint, specify an appropriate value under Temperature.Nominal.

ThermoPower

Off/On

On

Turns the power supply for the thermostat on or off.  If set to Off, it is not possible to start runs.

This also applies if a fraction collector  (G1364) is used.

Temperature.Nominal

4°C/40°C

20°C

Temperature set point for the air stream and the temperature control.

If a nominal value is specified, the TempCtrl and ThermoPower properties are automatically set to On.

Temperature.Value

4°C/40°C

n/a

Current temperature of the air stream. This temperature value is used as the setpoint for the temperature control, if TempCtrl = On, but no Temperature.Nominal value is specified.

Also refer to TempCtrl and Temperature.Nominal.

 

Also, see  Agilent HPLC System 1100/1200: Wellplate Sampler (G1367).

Diagnostic Commands

 

Command

Description

GripperHome

Moves the gripper to home position.

GripperPick

Picks the vial at the current position.

GripperPut

Positions the vial at the current position.

GripperToPosition

Moves the gripper to the specified position.

NeedleIntoSeat
NeedleIntoVial
NeedleToPosition

(For the G1329 autosampler, the commands have no functionality.)

NeedleUp

Moves the needle to its topmost position.

PlungerDraw

Draws the plunger to load sample into the sample loop.

PlungerHome

Moves the plunger back to its home position.

VialToSeat

Positions the vial in the seat.

VialToTray

Positions the vial in the tray.

ValveToBypass

Switches the sample loop into the bypass position.

ValveToMainpass

Switches the sample loop into the inject position.

 

For more information about the diagnostic functions for the autosampler, refer to:

 Agilent 1100 HPLC System Diagnostics: Injector Steps

 Agilent 1100 HPLC System Diagnostics: Gripper Verification

 

The Administrator Help section provides installation instructions for the 1100 and 1200 HPLC systems; refer to Hardware Installation:

 Agilent 1100 HPLC System: Installation

 Agilent 1200 HPLC System: Installation