Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: Autosampler (G1329)
In addition to (or different from) the standard autosampler commands (see Commands for Controlling Dionex Devices Dionex Autosamplers) and the general commands for the 1100 and 1200 HPLC systems (see Agilent 1100/1200 HPLC System: General), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
Min |
Max |
Default |
Description |
DispSpeed |
10 µl/min or |
1000 µl/min |
200 µl/min |
Speed for dispensing the sample |
DrawSpeed |
10 µl/min or |
1000 µl/min |
200 µl/min |
Speed for drawing the sample |
Position |
0 |
(Tray dependent) |
n/a |
Sample vial number |
SampleHeight |
-2.5 mm |
35 mm |
0.0 mm |
Needle draw/eject position offset |
Syringe |
100 µl |
900 µl |
as configured |
Syringe volume [µl] |
0.0 µl |
8000.0 µl |
5.0 µl |
Injection volume |
|
WashVial |
0 |
Greatest sample position (tray dependent) |
0 |
Wash vial number |
Command |
Description |
Injects sample into the loop. |
|
Draws sample from a vial into the syringe. |
|
InitiateChangeGripper |
Initiates the gripper change by moving the gripper to the front. |
InitiateChangeNeedle |
Initiates the needle change by removing the safety covering and moving the needle carrier to an easily accessible mounting position. In addition, the needle is raised to the uppermost position. |
InitiateChangePiston |
Initiates the piston change by releasing the spring on the proportioning unit. |
Performs an injection (see note below). |
|
Mix |
Mixes the sample using the syringe. |
Repeat/EndRepeat |
Repeats the enclosed sample preparation steps. |
TerminateChangeGripper |
Terminates the gripper change by returning the gripper to the Home position. |
TerminateChangeNeedle |
Terminates the needle changes by returning the needle to the Home position. |
TerminateChangePiston |
Terminates the piston change by tensing the spring on the proportioning unit. |
Valve |
Switches the injection valve. |
Wait |
Waits a given time during the injection program. |
Parameters of the commands
Command |
Values |
Description |
Dispense, |
SampleVial (current sample vial) |
Position of the vial for dispense |
Dispense/ Mix, |
10 (or 90 µl) to 1000 µl/min |
Speed for dispense or mix |
Draw, |
SampleVial (current sample vial) |
Position of the vial for draw |
Draw, |
10 (or 90) to 1000 µl/min |
Speed for draw |
Draw/Dispense, |
-2.5 to 35.0 mm |
Needle draw/eject position offset for draw or dispense |
Draw/ Dispense/ Mix, |
0.0 to 100.0 µl (or 900.0 µl) |
Volume for draw, dispense, or mix |
Needle, |
SampleVial (current sample vial) |
Position to which the needle is moved |
Tips:
The Inject command sends a start impulse to all modules. Then the analysis will start with the following steps:
The detector performs an Autozero if the program file contains the corresponding command at the time t=0.000 min.
An injection/sample preparation program is executed if the program contains the corresponding commands at the t=0.000 min.
The injection will take place and the run will start.
Three different injection programs are available:
Standard (injection volume from the respective position) |
WashVial=None and no sample preparation commands |
Standard (with needle wash) |
WashVial=None and no sample preparation commands |
User-defined injection program |
If sample preparation commands are issued |
Autosampler/Thermostat
Property |
Min/Max |
Default |
Description |
AutoActivateSetpoint |
Off/On |
On |
If set to On, temperature control is automatically enabled right after power-up. |
TempCtrl |
Off/On |
Off |
Turns temperature control on or off. If set to On, temperature control is activated and the current temperature as shown under Temperature.Value is used as the temperature setpoint. To define a different setpoint, specify an appropriate value under Temperature.Nominal. |
ThermoPower |
Off/On |
On |
Turns the power supply for the thermostat on or off. If set to Off, it is not possible to start runs. This also applies if a fraction collector (G1364) is used. |
Temperature.Nominal |
4°C/40°C |
20°C |
Temperature set point for the air stream and the temperature control. If a nominal value is specified, the TempCtrl and ThermoPower properties are automatically set to On. |
Temperature.Value |
4°C/40°C |
n/a |
Current temperature of the air stream. This temperature value is used as the setpoint for the temperature control, if TempCtrl = On, but no Temperature.Nominal value is specified. Also refer to TempCtrl and Temperature.Nominal. |
Also, see Agilent HPLC System 1100/1200: Wellplate Sampler (G1367).
Diagnostic Commands
Command |
Description |
GripperHome |
Moves the gripper to home position. |
GripperPick |
Picks the vial at the current position. |
GripperPut |
Positions the vial at the current position. |
GripperToPosition |
Moves the gripper to the specified position. |
NeedleIntoSeat |
(For the G1329 autosampler, the commands have no functionality.) |
NeedleUp |
Moves the needle to its topmost position. |
PlungerDraw |
Draws the plunger to load sample into the sample loop. |
PlungerHome |
Moves the plunger back to its home position. |
VialToSeat |
Positions the vial in the seat. |
VialToTray |
Positions the vial in the tray. |
ValveToBypass |
Switches the sample loop into the bypass position. |
ValveToMainpass |
Switches the sample loop into the inject position. |
For more information about the diagnostic functions for the autosampler, refer to:
Agilent 1100 HPLC System Diagnostics: Injector Steps
Agilent 1100 HPLC System Diagnostics: Gripper Verification