Thermo Finnigan Surveyor MSQ Mass Spectrometer: Installation
Device Type: |
Single quadruple Mass Spectrometer |
Device Driver: |
Thermo Scientific: Thermo Finnigan MSQ MS (tested for firmware version 1.0) |
Supported Hardware Options: |
No special options available. |
What is required?
License: |
Timebase Class 1 + MS Control |
Connection: |
|
Hardware |
A free RS-232 port is required for control. |
MS_FULL_SCAN.pan |
Hardware Installation
Device Connection: |
Data acquisition cannot be started from Xcalibur software. Therefore, you must connect the Inject-Output-Signal of the LC Autosampler to the Thermo Finnigan MSQ user interface IN1 remote input. (Use the black and red wires of the 3-wire cable that is shipped with the MSQ mass spectrometer.) |
Software Installation
For information about the software installation
required for data acquisition, refer to How
to: Installing and Configuring Mass Spectrometers Installing MS Components.
Install the Device
Driver in the Chromeleon Server Configuration. (For more information
about the program, refer to Software
Installation and Communication The
Server Configuration
Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the mass spectrometer to the timebase. Select Add Device on the Edit or context menu, and then select Thermo Scientific from the Manufacturers list box and Thermo Finnigan MSQ MS from the Devices list box.
The following settings are required:
On the General tab page, select the port to which the mass spectrometer is connected. In addition, disable the Demo Mode.
Do not change the settings, that is, the number of channels installed, on the Installed Channels tab page.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the Thermo Finnigan MSQ Mass Spectrometer, connect one of the following Control Panels in the Chromeleon client to the corresponding timebase: MS_FULL_SCAN.pan MS_LC_PDA.pan, MS_LC_UV.pan, or MS_SIM_MODE.pan.
Application
For information about the commands and properties
supported by the MSQ mass spectrometer, refer to Commands
and Tips for Thermo Scientific
Devices Thermo Finnigan Surveyor MSQ Mass
Spectrometer in the User Help
section.
Troubleshooting
For information about troubleshooting, refer
to Commands and Tips for Thermo Scientific Devices
Thermo Finnigan MSQ: Troubleshooting
in the User Help section.
Further Information
Tip:
For more installation details, refer to the manual for the instrument.