Commands and Tips for Thermo Scientific Devices
Thermo Finnigan Surveyor MSQ Mass Spectrometer

TIC, SIM, and MS Channel Properties and Commands

These channels are standard data channels with the following properties (please note that the display Filter level determines which properties are displayed):

Property

Min

Max

 

Description

AcqOn/Off

-

-

MS_n only

Starts and stops data acquisition.

MinMass

0.00 m/z

2000.00

Read-only

Minimum mass recorded in this trace. (SIM channels only)

MaxMass

0.00 m/z

2000.00

Read-only

Maximum mass recorded in this trace. (SIM channels only)

Polarity

Negative

Positive

Read-only

Polarity

SourceVoltage

0 V

200 V

Read-only

Maximum voltage.

FilterIndex

0: NoFilter

n (dep. on installation):

9 (TIC/MS)

41 (SIM)

Read-only
(for TICF and SIM channels)

Extraction filter. The filter corresponds to the filters defined in the MS method. For 0 = NoFilter, the TIC channel is used.

Filter

n/a

n/a

Read-only

Filter string as used in Thermo Finnigan MSQ method.

BasePeakMode

No

Yes

MS only

Extracts Base Peak chromatograms during data acquisition

Delta

-

-

Read-only

Signal change (difference between the current value and the value one second before).

 

 Tip:

Support for data compression is not available. Thus, the Step, MaxAutoStop, and Average commands cannot be set. Manual data acquisition is not possible, either. In addition, acquisition is started at the time 0.000 and stopped at the end of the program for all channels that deliver data according to an MSQ method file. These channels do not appear in the Program Wizard or on the control panel.

 

Device Properties and Commands

Property

Min

Max

Default

Description

Ready

Ready, Busy

 

Read-only

Indicates whether the spectrometer is busy or idle.

Operation

On, Off, Standby

 

Off

Indicates whether the spectrometer is busy or idle.

Status

n/a

 

Read-only

Indicates the spectrometer's operational status.

Range

1
(for exp(1) Counts)

10
(for exp(10) Counts)

 

Scaling range used for online signal plot during data acquisition. (Can be set only before data acquisition starts.)

Smoothing

None, Gaussian, Boxcar

None

 

Smoothing algorithm.

Note: This parameter is used by Xcalibur raw data extraction during online and post-run channel extraction.

SmoothingPoints

3, 5, 7, 9,
11, 13, 15

3

 

Number of data points used for smoothing.

Note: This parameter is used by Xcalibur raw data extraction during online and post-run channel extraction.

HardwareVersion

n/a

 

Read-only

The mass spectrometer's hardware version.

Note: This value is available only after the first raw data acquisition.

FirmwareVersion

n/a

 

Read-only

The spectrometer's firmware version.

Note: This value is available only after the first raw data acquisition.

AbortAcquisition

 

 

 

Aborts the current Xcalibur acquisition and resets the MS. Use this command in a trigger command, for example:

Trigger EndNow <condition>

MS.AbortAcquisition; ends MS acq

End; ends PGM and UV acq

EndTrigger

 

Command

Description

Connect

Connects the device

Disconnect

Disconnects the device

Reset

Resets the device to the default values

 

For troubleshooting information, refer to  Thermo Finnigan Surveyor MSQ: Troubleshooting.