DGP-3600
and LPG-3400 (RS, SD or BM, not
XRS) pumps only:
Specify
the settings for pump stroke synchronization when one of the pumps
listed above is operated with an OAS-3x00TXRS
autosampler. In addition,
you can specify a shutdown action for the Corona detector, in
case the pump flow decreases to 0.
OAS-3x00TXRS
Sync – Select this check box
to enable pump stroke synchronization for the pump. When using a dual DGP-3600RS/SD/BM
pump, you can specify whether synchronization is to be performed
for the left or right pump unit.
Clear
the check box to disable pump stroke synchronization with the
autosampler. For more information about how to use the pump with
an OAS-3x00TXRS, refer to UltiMate 3000 Series OAS-3x00TXRS Autosampler:
Installation.
Emergency
Shutdown (Coulochem/ECD-RS) – Selecting
this check box enables a Coulochem
or ECD-RS detector to automatically start a Shutdown procedure
when the pump flow becomes 0. This option is enabled by default.
When
using a dual DGP-3600RS/SD/BM pump, you can specify whether synchronization
is to be performed for the left or right pump unit.
Emergency Shutdown (Corona)
- Selecting this check box enables a HPLC pump to automatically
start a Shutdown procedure when the gas flow becomes 0. This option
is enabled by default. |