Installing Dionex Devices
UltiMate 3000 Series OAS-3x00TXRS Autosampler
Device Type: |
|
Device Driver: |
Dionex HPLC: UltiMate 3000 OAS-3X00TXRS |
Hardware
Options |
For detailed information about the available configurations, refer to the operating instructions for the autosampler. |
Restrictions: |
|
What is required?
Hardware Installation
To connect the autosampler with the Chromeleon Server PC There are two possible types of communication between the Chromeleon Server PC and the autosampler: Ethernet (LAN) for communication via TCP/IP or serial via RS-232 communication.
By default, the autosampler is configured for communication via TCP/IP. The default IP address for the OAS-3x00TXRS is as follows:
To establish LAN communication, proceed as follows: Establish the LAN connection between autosampler and Server PC by connecting the instrument to an Ethernet switch or hub using the standard Ethernet LAN cable shipped with the instrument. Alternatively, you can use a Standard crossover cable to connect the autosampler directly to the Server PC. Tip: For better reliability, Thermo Fisher Scientific recommends connecting the autosampler to a separate device LAN and not to the Office LAN. Note that an additional network card is required when using a device LAN. For information about how to set up the TCP/IP connection, also refer to Connect Modules via TCP/IP (LAN) and the related subtopics.
As an option, serial communication via RS-232 can be set up for the autosampler. However, LAN communication is the recommended type of communication. To establish serial communication, proceed as follows:
|
|||||||||||||||||
Device Settings: |
If required, communication settings for the autosampler can be changed by using the autosampler's Control Terminal. To set the communication protocol Switch the autosampler from Ethernet (LAN) to serial (RS-232) communication or vice versa, if required. The current setting (SER or LAN) is shown during boot-up.
To set the IP address, subnet masks, or gateway address Depending on the LAN environment and number of autosamplers used, you may want to change the IP address or related settings.
Note: If you are controlling more than one OAS-3x00TXRS autosampler via LAN from the Chromeleon Server PC, each autosampler must have a unique IP address. The IP address for the OAS-3x00TXRS is 172.16.1.200, where the fourth identifier (.200) is a unique integer value for each autosampler connected to the PC. For detailed information about how to use the Control Terminal, refer to the OAS-3x00TXRS operating manual. |
||||||||||||||||
Hardware Synchronization: |
Typically, the autosampler is operated with one of the following UltiMate 3000 devices:
Connect the autosampler with an UltiMate 3000 RS, SD, BM or XRS pump and/or Corona or Coulochem detector, as described in topic UltiMate 3000 Series OAS-3x00TXRS Autosampler: Connecting the Autosampler. After successful device connection, you have to configure synchronization settings on the Control Terminal of the autosampler to enable:
For details, see sections Inject Response and Pump Stroke Synchronization below.
Inject Response To ensure that the inject response signal is transmitted properly, the OUT signal settings for the autosampler have to be configured as follows:
Pump Stroke Synchronization In pump stroke synchronization, the pump strokes are synchronized with the injection command of the autosampler. Synchronization ensures that all injections are performed at the same phase of the pump cycle; this enhances the retention time precision for gradient applications. To enable pump stroke synchronization, you have to define the pump unit used for synchronization on the Open Autosampler OAS-3x00TXRS page (option Pump Link (LPG only)) in the Chromeleon Server Configuration. Exception: When using a LPG-3400XRS pump, set Pump Link (LPG only) to None. Then change to the autosampler's control terminal to specify the synchronization signal:
To disable pump stroke synchronization: If the pump does not support pump stroke synchronization or no synchronization is needed, you can disable pump stroke synchronization as follows:
|
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
First, configure the following pump settings:
If the autosampler is operated with a UltiMate 3000 RS (not XRS), SD or BM pump: On the Inputs page of the pump, select the pump input to enable synchronization of the inject response signal.
Then specify
the required settings for pump stroke synchronization:
If the autosampler is operated with a DGP-3600
or LPG-3400 (RS, SD or BM) pump: On the Relays
page of the pump, in the Pump Links
group, select the OAS-3x00TXRS
check box to enable pump stroke synchronization. When using a DGP-3600
dual pump, select whether the left or right pump unit should be used
for pump stroke synchronization. Note that when synchronization is
enabled, the relays are controlled by the pump and are no longer available
in Chromeleon for use for different purposes. Disable any currently
selected relays, if necessary.
If no pump stroke synchronization is required, clear the OAS-3x00TXRS
check box.
Notes:
Observe the following when operating a dual DGP-36000RS, SD, BM pump:
It is not possible to enable pump stroke synchronization for both pump units at the same time.
If pump stroke synchronization is enabled for one of the pump units, the Emergency Shutdown option can also only be selected for one pump unit. Clear the Emergency Shutdown check box for the second pump unit in this case.
After configuration of the pump settings, proceed to the autosampler configuration. On the Open Autosampler OAS-3x00TXRS configuration page, configure the settings as needed. In particular, specify:
whether instrument communication with the Chromeleon Controller PC is via COM port (serial) or via TCP/IP (LAN),
how the inject response signal is transmitted between the autosampler and its connected devices (pump or detector), and
if pump stroke synchronization is needed or not. For information about the settings on the page, click Help or press F1.
Further Information
For detailed installation instructions, refer to the operating instructions for the OAS-3x00TXRS.
For troubleshooting information, refer to UltiMate 3000 Series OAS-3x00TXRS Autosampler: Troubleshooting.
For information about how to connect the autosampler with the modules of the UltiMate 3000 series, refer to UltiMate 3000 Series OAS-3x00TXRS Autosampler: Connecting the Autosampler.
For troubleshooting information, refer to Troubleshooting for USB and TCP/IP Connections.