Installing Dionex Devices
UltiMate 3000 Series NCP-3200RS and NCS-3500RS: Installation
Device Type: |
Nano/Cap pump (NCP-3200RS) and Nano/Cap system (NCS-3500RS) for UHPLC applications |
Device Driver: |
Dionex HPLC: UltiMate 3000
|
Hardware
Options |
NCP-3200RS:
NCS-3500RS:
For information about the available configurations, refer to the operating instructions for the NCP/NCS. |
Restrictions: |
Defining of gradient curves for the Nano flow is only supported with firmware version 1.32 or later. |
Backward Compatibility |
ProFlow Flowmeters NC pumps equipped with ProFlow flowmeters are supported by Chromeleon 6 versions starting with 6.80 SR15b. NCS or NCP driver configurations created with Chromeleon versions before 6.80 SR15b, however, cannot immediately control a ProFlow flowmeter after an upgrade to Chromeleon 6.80 SR15b. To control an NC pump with ProFlow flowmeter, remove the old NCS or NCP driver instance from your timebase and then re-install the driver in Chromeleon 6.80 SR15b or later. |
What is required?
License: |
Timebase Class 1 |
Connection: |
A free USB port and the following cables are required:
|
To control the instrument, use |
Hardware Installation
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
On each configuration tab page, select the desired settings. For information about the settings on a page, click Help or press F1.
Note:
After a software upgrade from CM 6.80 SR9 to a higher Chromeleon version, existing NCS-3500RS drivers may display the following default column name on tab page Columns (NCS-3500RS-System): Column1...Column4. This name, however, is inconsistent with the column names used on the Dionex default panel or panel tabset for the NCS-3500RS (ColumnA...ColumnD). As a result, an error message in the Chromeleon Client software may occur.
To avoid errors caused by inconsistent names, it is recommended to change the column name on the Columns tab page in the Server Configuration to ColumnA...ColumnD.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For troubleshooting information, refer to Troubleshooting for USB and TCP/IP Connections.
For an overview of the different Dionex devices, refer to Installing Dionex Devices.