Installing Dionex Devices
Dionex ESA Corona (plus/ultra/ultra RS) CAD: Installation

 

Device Type:

Charged Aerosol Detector

Device Driver:

Dionex ESA Modules: Corona Charged Aerosol Detector

Hardware Options
Supported:

The driver supports the following Corona models:

  • Corona CAD

  • Corona plus (same as Corona CAD but with thermal nebulizer)

  • Corona ultra

  • Corona ultra RS (same as Corona ultra but with flow control, waste level monitoring, signal linearization and an expanded filter set)

Restrictions:

The following detector functions are not supported:

  • Inputs and analog outputs

  • Loops in timeline program

  • Self test

  • Firmware download

 

What is required?

 Note:  

If you run Agilent LCs controlled by Empower, you require the Corona DII connection kit (6081.2100) for connecting the Corona detector. Installation instructions are provided with the connection kit.

License:

Timebase Class 1

Connection:

USB Connection

A free USB port and the following cable is required:

  • USB cable (Dionex part no. 6911.0002) to connect the device to the server PC via the USB port.

  • A 6-wire cable (Dionex part no. 70-9419) is required to connect the Waste Bottle/Waste Level Detector Assembly (Dionex part no. 70-9363R) to the Corona ultra RS.

Pump Connection

If the detector is operated with an UltiMate 3000 pump, the following additional signal cables are required:

- UltiMate 3000 LPG-3400XRS pump:

  • Synchronization cable (Dionex part no. 6043.0003). The cable is shipped with the detector.

    Detailed instructions for connecting the Corona detector with an LPG-3400XR pump are also provided in the Operating Instructions of the pump.

 

- UltiMate 3000RS/SD/BM/BX pumps:

  • 6-pin Mini-DIN signal cable (Dionex part no. 6000.1004)

  • Required pump firmware version: 2.20 or higher

- All other UltiMate 3000 pumps:

  • Shutdown cable (Dionex part no. 6035.0004).

  • Required pump firmware version: 3.05 or higher

    For details on installing the cables, refer to

     ESA Corona CAD: Interfacing the detector with an UltiMate 3000 Pump

     Dionex/ESA Interface Cable

 

For details on cable connections when operating the Corona detector with an UltiMate 3000 pump and an OAS-3x00TXRS autosampler, refer to:

UltiMate 3000 Series OAS-3x00TXRS Autosampler: Connecting the Autosampler.

Limitations:

If multiple Corona detectors are installed on a single PC, communication problems may occur in the following situations:

  • when switching off one of the detectors,

  • when connecting an additional detector with the PC.

To resolve the communication problems, you have to re-assign the USB address to the detector. To do so, on the General tab page, click Browse in order to load the correct USB address again.

Control Panel:

To control the detector, use

  • panel tabsets (refer to How to:  Controlling Devices from the Panel Tabset)

    OR

  • Corona CAD.pan from the
    Dionex Templates\Panels\Dionex_LC
    folder in the Browser

 

Hardware Installation

Safety Connection:

 Caution:

Dionex recommends connecting the PUMP OFF output contact of the detector to the pump, so that, if the pressure of the nebulizing gas falls below 50% of the required gas volume, the signal may be used to shut the pump off so that the detector is not flooded with mobile phase.

Device Connection:

First, install the Chromeleon software, and then connect the instrument to the USB port on the Chromeleon server PC.  

  1. The Windows Found New Hardware Wizard starts.

  2. Complete the wizard, selecting these options:

    When asked if Windows can search for software, select No, not this time.

    When asked to choose an installation type, select Install from a list or specific location (Advanced).

    When asked to specify the search options, select Include this location in the search.

  1. Click the Browse button and specify the following search location: C:\Chromel\bin\DDK\V1\Drivers\Dionex\Corona.

  2. When the wizard is finished, the device will be listed in the Device Manager under Dionex USB I/O Devices.

Also refer to Connecting Dionex Devices via USB.

 

Software Installation

  1. In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.

  2. On each configuration tab page, select the desired settings. For information about the settings on a page, click Help or press F1.

Further Information

For detailed installation instructions, refer to the operating instructions for the instrument.

For troubleshooting information, refer to Troubleshooting for USB and TCP/IP Connections.

For an overview of the different ESA devices, refer to ESA Overview.