Diagnostics (UltiMate 3000 Series)
Diagnostic Tests for UltiMate 3000 Instruments
The table below lists the UltiMate 3000 instruments and the various diagnostics tests that are available for an instrument. Some tests, however, may not be available for or applicable to specific instrument types of the UltiMate 3000 series. For information about the tests that are provided for an instrument type, refer to the Chromeleon Client program: click the Control menu and then select Diagnostics to display a list of all diagnostics tests that are available for the UltiMate 3000 instrument(s) installed in a timebase.
Device |
Run the... |
To check the ... |
Pumps*
|
pump for leakage. |
|
valves and seals for leakage. |
||
pulsation, compression. |
||
permeability of the outlet/mixer frits. |
||
degasser vacuum. |
||
NCP-3200RS/NCS-3500RS |
|
|
NC pump |
offset of the pressure transducers. |
|
|
the NC pump for leakage. |
|
|
Use this test to determine the viscosity of a solvent that you want to use with the NC pump but for which the viscosity is not yet available in Chromeleon. The test also checks whether the pump works correctly and delivers the exact flow rate. |
|
|
(ProFlow flowmeter only) Use this test to set the flow sensor and pressure transducer to zero. |
|
|
(ProFlow flowmeter only) Use this test to calibrate solvents to be used with a ProFlow flowmeter. |
|
Loading pump |
see the tests listed under Pumps (further up in this list) |
|
|
|
|
Detector VWD-3x00 |
UV lamp wavelength calibration. |
|
noise and dark current signal. |
||
correct functioning of the filter motor. |
||
correct functioning of the grating motor. |
||
wavelength accuracy. |
||
lamp intensity. |
||
correct functioning of the filter for suppressing high-order light proportion. |
||
|
|
|
DAD-3x00/MWD-3x00 |
wavelength accuracy. |
|
|
lamp
intensity (with or w/o flow cell) |
|
|
dark current signal (with or w/o flow cell). |
|
(RS versions only) |
slit mechanism. |
|
|
|
|
Autosampler WPS-3000SL/RS/TXRS |
the autosampler motors, sensors and injection valve. |
|
autosampler or HPLC system for leakage. |
||
|
|
|
Column Compartment TCC-3x00 |
gas sensor in the column compartment. |
|
thermal output of the thermo unit, fans and thermal connections. Note: This test is not available for TCC-3000RS/SD column compartments. |
||
|
|
|
Flow Manager |
components of the flow control system (pressure sensor, flow control valve, flow sensor, flow splitter). |
|
gas sensor in the column compartment. |
||
|
thermal output of the thermo unit, fans and thermal connections. |
*To run diagnostic tests for the UltiMate 3000 pumps, verify that the appropriate firmware version is installed.
For general information about the diagnostic tests, refer to Diagnostics (UltiMate 3000 Series).