Diagnostics (UltiMate 3000 Series)
Pressure Transducer Test (NCP-3200RS/NCS-3500RS)
The NC pump generates the flow rate by applying a corresponding pressure drop at the flow selector. To ensure an accurate result, the pressure transducers must be synchronized very well. Run the Pressure Transducer Test to verify that the pressure transducer offset is still correct. If it is not, calibrate the pressure transducers.
In addition, if you observe a shift in peaks or if the gradients are not correct, also consider checking the pressure transducer offset.
How to
On the Control menu, select Diagnostics, and then select Pressure Transducer Test. A wizard guides you through the procedure. After calibration, the new value is stored automatically as the current value.