Installing and Controlling Thermo Scientific Devices
Thermo Scientific Surveyor Plus Autosampler
Device Type: |
|
Device Driver: |
Thermo Scientific: Surveyor Plus Autosampler |
|
The autosampler was tested with firmware version 2.15. Note: The "Thermo Surveyor AS" model can be controlled by the Thermo Surveyor Plus Autosampler driver if firmware version 2.15 or later is installed. |
Restrictions:
|
The following options are not supported:
|
What is required?
License: |
Timebase Class 1 |
Connection: |
Note: The System Interconnect Cable supplied by Thermo Scientific for all Surveyor Plus modules is not required when running the system under Chromeleon control. Disconnect the cable if attached to the system and verify that the Start and Stop inputs are not connected when running the Surveyor Plus system under Chromeleon control. Otherwise, Chromeleon will not be able to perform synchronization between modules. |
Windows Firewall Settings: |
For the Windows Firewall settings that are required for Windows XP, Windows Vista and Windows 7, respectively, refer to Chromeleon and Windows Firewall. |
To control the instrument, use the Thermo_Surveyor_Plus_Autosampler.pan from the |
Hardware Installation
Software Installation
In the Chromeleon Server Configuration, add the Driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver to the timebase, the Configuration dialog or wizard opens automatically.
On the General tab page specify the following settings:
In the Module ID box, type or select the unit ID value for the instrument. The module ID must correspond with the value set by the rotary switches at the rear panel of the autosampler. If you need to change the switch settings, see instruction in section Device Settings above.
Note: The module ID is required to specify the network IP address of the autosampler. The IP address is in the format 172.16.x.111, whereby 'x' must correspond with the module ID of the autosampler.
Click the arrow next to the Inject Sync box to select the COM port to which the Dionex synchronization cable is connected with the PC.
On the Signals tab page, the following check boxes must be selected, if the autosampler is operated with a Surveyor Plus pump:
Pump Ready Input Active High,
Inject Hold Release Input Active High and
Inject Output Active High.
On the Configuration tab page, specify settings as required.
For information about the settings on a page, click the Help button or press F1.
Tip:
The only supported configuration is a Surveyor Plus pump in combination with a Surveyor Plus AS. It is not possible to control either module alone, or in combination with Accela modules.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
Troubleshooting for USB and TCP/IP Connections
Thermo Scientific - for an overview of the different Thermo Scientific/Thermo Finnigan/ThermoQuest/TSP instruments that can be controlled by Chromeleon.