Installing and Controlling Thermo Scientific Devices
Thermo Scientific Surveyor Plus Autosampler

Device Type:

Autosampler

Device Driver:

Thermo Scientific: Surveyor Plus Autosampler

 

The autosampler was tested with firmware version 2.15.

Note:

The "Thermo Surveyor AS" model can be controlled by the Thermo Surveyor Plus Autosampler driver if firmware version 2.15 or later is installed.

Restrictions:

 

The following options are not supported:

  • Sample preparation

  • Timed events

 

What is required?

License:

Timebase Class 1

Connection:

  • A free Ethernet port

  • A free COM port

  • Ethernet LAN cable, which is shipped from Thermo Scientific with the autosampler

Note:

The System Interconnect Cable supplied by Thermo Scientific for all Surveyor Plus modules is not required when running the system under Chromeleon control.

Disconnect the cable if attached to the system and verify that the Start and Stop inputs are not connected when running the Surveyor Plus system under Chromeleon control. Otherwise, Chromeleon will not be able to perform synchronization between modules.

Windows Firewall Settings:

For the Windows Firewall settings that are required for Windows XP, Windows Vista and Windows 7, respectively, refer to Chromeleon and Windows Firewall.

Control Panel:

To control the instrument, use the Thermo_Surveyor_Plus_Autosampler.pan from the
Dionex Templates\Panels\LC_Other
folder in the Browser.

 

Hardware Installation

Device Connection:

  1. Connect the autosampler to an Ethernet switch or hub using the Ethernet LAN cable shipped with the instrument.

  2. Configure the TCP/IP network settings: Enter the Subnet Mask (255.255.0.0) and IP address (172.16.yy.zz)  to be used. Note that the addresses 172.16.0.0 and 172.16.255.255 are reserved for special purposes and therefore cannot be used.

For information about how to setup the TCP/IP connection, also refer to  Connecting Devices via TCP/IP (LAN) and the related subtopics.

Device Settings:

If necessary, use the rotary switches located on the rear panel to change the module ID (and thus the IP address) of the autosampler:

The range of values is from 01–99. The default value is 01. The value 00 is reserved for special service settings.

After changing the rotary switch settings, the autosampler must be powered OFF and ON again to allow the changes to take effect.

Note:

When connecting more than one Surveyor Plus autosampler via Ethernet, make sure to configure each autosampler with a different module ID.

For detailed information, refer to the hardware manual for the autosampler.

Connecting the Synchronization Cable:

The Dionex synchronization cable is an optional accessory that allows improved synchronization between the autosampler and Chromeleon. The cable has three connectors for connecting the Surveyor Plus autosampler with the Surveyor Plus LC pump, and the COM port on the Instrument Controller PC.

Connect the cable as follows:  

  1. Plug the autosampler connector (marked "A/S") of the synchronization cable into the left-most 8-pin socket at the rear panel of the autosampler (see figure below).

  2. Plug the pump connector (marked "Pump" of the synchronization cable into the left-most 8-pin socket at the rear panel of the Surveyor Plus pump.

  3. Attach the 9-pin D-Sub (female) connector of the synchronization cable into the serial port of the PC. If the PC does not provide a serial port, use a serial interface card (Meilhaus)instead. For more information about Meilhaus cards available from Thermo Scientific Dionex , refer to Installing the Serial RS-232 Interface Card (Meilhaus).

    Note:

    Connection via UCI COM port is not possible.

 

Software Installation

  1. In the Chromeleon Server Configuration, add the Driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver to the timebase, the Configuration dialog or wizard opens automatically.

  2. On the General tab page specify the following settings:

  3. On the Signals tab page, the following check boxes must be selected, if the autosampler is operated with a Surveyor Plus pump:

  4. On the Configuration tab page, specify settings as required.

For information about the settings on a page, click the Help button or press F1.

Tip:

The only supported configuration is a Surveyor Plus pump in combination with a Surveyor Plus AS. It is not possible to control either module alone, or in combination with Accela modules.

 

Further Information

For detailed installation instructions, refer to the operating instructions for the instrument.

Troubleshooting for USB and TCP/IP Connections

 Thermo Scientific - for an overview of the different Thermo Scientific/Thermo Finnigan/ThermoQuest/TSP instruments that can be controlled by Chromeleon.