Connecting Devices to the Chromeleon Server
Troubleshooting for USB or TCP/IP Connections
The following table lists the most important error messages that may appear in the Audit Trail if the connection between Chromeleon and the device is not successful. In addition, possible causes and the corresponding remedial actions are suggested. In this case, the messages refer to the P680 pump.
Tips:
Please keep in mind that, for up to 30 seconds after the server configuration is changed, messages referring to the previous configuration may be displayed. These messages can be ignored. Thus, always note the device name referred to in the message; for example, P680@...
The error messages below may also read either P680@ + serial number (= P680@USB-01610103), or P680@ + IP address (= P680@192.168.50.1), depending on the connection type (USB or TCP/IP) specified for the instrument in the Server Configuration program. (For examples for the USB connection, refer to the second table below.)
Message |
Probable Cause |
Remedial Action |
Error opening P680 - Connection refused. |
Another server of the Chromeleon already uses the indicated pump. |
Check that the correct pump was selected or terminate the pump's communication with the other Chromeleon server. |
The existing connection is not yet cleared completely. |
Wait for 30 seconds and repeat the procedure. |
|
The pump's subnet mask and gateway address configuration is incorrect. |
Make sure that the correct parameters are specified. |
|
A gateway may maintain an unintended connection to the pump as soon as the Chromeleon server is stopped in a non-controlled way. |
Restart the pump. |
|
Error opening P680 - Connection timed out. |
The Chromeleon server failed to connect to the indicated pump. |
Verify that the pump is turned on and that the network connection is working. |
The network connection is interrupted during firmware download. |
Check the network connection. |
|
The network connection is interrupted during firmware download. |
Check the mains connection of the pump. |
|
P680 - Connection reset by peer. |
The power supply to the pump is interrupted for a short period. |
Check the mains connection of the pump. |
P680 - Host not found. |
Network problems: No connection to the pump. |
Check the network settings for the |
P680 - Permission denied. |
||
P680 - Network is unreachable. |
||
P680 - No route to host. |
||
Error issuing control request to P680 |
The Chromeleon server cannot connect to the specified pump. |
Check the USB and LAN connection, respectively. Check the mains connection of the pump. Remove the pump specified in the message from the server configuration or else select a different pump from the list of available pumps in the server configuration (via Properties/ Browse). |
The following error messages refer to the USB connection of a P680 pump. Similar messages may also appear for the following devices: UVD 170U, UVD 340U, LPG-3x00, FLM-3x00, and TCC-100:
Message |
Probable Cause |
Remedial Action |
Error reading from P680@USB-00- |
There is a transmission error between the pump and the Chromeleon server. |
Check the USB connection. The connection length must not exceed 5 m to the next hub. The overall connection length, including the hub connections, must not exceed 30 m. Remove any extension cables even if both rules are met. Replace any defective USB cable or hub. |
Error reading from P680@USB-00- |
The connection between the pump and Chromeleon server is interrupted. The power supply to the pump is interrupted |
Check the USB connection. Check the mains connection of the pump. |
P680@USB-00- |
The USB connection between the pump and the Chromeleon server is interrupted. The power supply to the pump is interrupted |
Check the USB connection. Check the mains connection of the pump. |
Error opening P680@USB-00- |
The USB connection between the pump and the Chromeleon server is interrupted. The power supply to the pump is interrupted |
Check the USB connection. Check the mains connection of the pump. |
For a list of all error messages that might appear in Chromeleon, see the Error Levels tab page in the Server Configuration program.