Installing and Controlling Third-Party Devices
Fisons AS800 GC Autosampler: Installation

Device Type:

GC Autosampler

Device Driver:

Fisons: FisonsAS800_Sampler or
ThermoFinnigan/Thermoquest/TSP: AS2000 GC Sampler
(for operating the autosampler in dual operation)

Supported Hardware Options:

 

No special options available.

 

What is required?

License:

Timebase Class 1

Connection:

RS-232 cable (For cable details, refer to  ThermoQuest//Fisons Cable.)

Hardware
Prerequisites:

  • An A/D converter is required for data acquisition. A UCI Universal Chromatography Interface including the appropriate signal cables may be used.

  • A free RS-232 port is required for control.

Software
Prerequisites:

No software prerequisites known.

Control Panel:

FISAS800.pan

Hardware Installation

Device Connection:

A special cable (Thermo Scientific part no. 8914.0148) is required to connect the autosampler to the PC (refer to  ThermoQuest/Fisons Cable).

Use a 37-pin connecting cable to connect the Fisons GC Autosamplers to the Fisons 8000 Gas Chromatograph. When the injection is completed, the GC autosampler sends the signal to the GC, which in turn sends it to Chromeleon via a relay. The pins 12 and 24 of the GC AUX connector are used. Connect the signal and GND lines of a remote input to these.

 

GC Fisons 8000   Remote Input
AUX Connector  (for example,
UCI-100)

(25-pin; Sub-D, female)  (9-pin, Sub-D, male)

12 ----------- Signal

24 ----------- GND (0V)

 

In the Server Configuration, specify the remote input that is used by the Fisons AS800 autosampler under Inject Port on the Sampler tab page!

Device Settings:

The following communications parameters must be used: 9600 Baud, 8 Data Bits, No Parity, 1 Stop Bit, Rts/Cts handshake.

Some Fisons AS800 GC samplers are set to a baud rate (by the manufacturer) that is not suitable for the Chromeleon device Device Drivers.

When turning on the autosampler, simultaneously press the Meth and Seq keys to check the baud rate setting. The following information will be displayed:

AS800 SET UP: b9600

ACT INJ SMP SYR MAN

The value on the top right is the baud rate setting. This information may not be available in previous firmware versions. If this is the case, press the EXIT key to leave the Setup menu.

If the baud rate is not set to 9600, move the cursor to the baud rate field using the left arrow key (<). Set the new baud rate, using the up and down arrow keys. Press the Exit key to leave the Setup menu.

Device Settings
(Dual Operation):

Only if you want to operate two autosamplers in dual operation mode:

On the autosampler display, set Sync to Twin (program page PP14).

 

Software Installation

Install the Device Driver in the Chromeleon Server Configuration. (For more information about the program, refer to Software Installation and Communication  The Server Configuration Program.)

  1. Start the Chromeleon server.

  2. Start the Server Configuration program.

  3. Add the autosampler to the timebase. Select Add Device on the Edit or context menu, and then select Fisons from the Manufacturers list box and FisonsAS800_Sampler (or ThermoFinnigan/Thermoquest/TSP: AS2000 GC Sampler) from the Devices list box.

  4. The following settings are required:

On the FisonsAS800_Sampler tab page, disable the Demo Mode.

On the Sampler tab page, select a free inject port.

The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.

  1. To control the Fisons autosampler connect the FISAS800.pan control panel in the Chromeleon client to the corresponding timebase.

Application

Note: The AS800 autosampler is identical in constructions to the ThermoQuest AS2000 autosampler. The User Help section provides more information about the AS800 Autosampler; refer to Commands and Tips for Third-Party Devices:

 ThermoQuest AS2000/Fisons AS800 Autosamplers

 ThermoQuest AS2000/Fisons AS800: Program Example

Troubleshooting

The User Help section provides tips for troubleshooting; refer to Commands and Tips for Third-Party Devices  ThermoQuest AS2000/Fisons AS800: Troubleshooting.

Further Information

 Tip:

For more installation details, refer to the manual for the instrument.

The Fisons AS800 autosampler corresponds to the ThermoQuest AS2000 autosampler. For more information, refer to  ThermoQuest AS2000 GC Autosampler.

For an overview of the different Fisons instruments for which device drivers are available in Chromeleon, refer to  Fisons.