Tab Dialog
Box:
Signals
The Signals tab page lists all signals or Channels that are available for the device. Which options are available depends on the device type.
Option |
Description |
|
|
Enabled |
Select a check box to enable the related signal or channel. Data can be recorded only for the selected signals (channels). If a check box is cleared, the signal (channel) will not be available in Chromeleon. |
|
|
Type |
Indicates the type of the signal. Also see information in section Tips for NC Pumps below. |
|
|
Name |
Indicates the signal name. To change a name, overwrite the name directly in the field. |
|
|
Unit |
(Not provided in all instruments) Specify the signal unit for the recorded data, for example, µS, nA, mAU, °C. |
|
|
Factor |
(Not provided in all instruments) Specify the factor for the selected signal (channel). |
It depends on the installed flowmeter type, which channels should be enabled for recording pump signals. The below described settings are set by default upon initial configuration of the instrument.
If the flowmeter type is changed later, however, verify that the channel settings for the pump are still set correctly. For example, if the flowmeter type was changed from Classic to ProFlow, make sure that the channels are set as described for the NC pump with ProFlow flowmeter above.
For more details about exchanging the flowmeter and recording pump signals, also refer to the Operating Instructions of the NC pump shipped with the instrument.
If an NC pump with ProFlow flowmeter is used, the following channels have to be selected:
LoadingPump_Pressure (NCS-3500RS only)
NC_Pump_Pressure
NC_Pump_Flow
NC_Pump_Flow_LeftBlk
NC_Pump_Flow_RightBlk
If an NC pump with Classic flowmeter is used, the following channels have to be selected:
LoadingPump_Pressure (NCS-3500RS only)
NC_Pump_Pressure
NC_Pump_Press_LeftBlk
NC_Pump_Press_RightBlk
For information about how to install the device, refer to:
ICS-Series Variable Wavelength Detector
UltiMate 3000 Series - NCP-3200RS/NCS-3500RS
UltiMate 3000 Series - Pumps (BM/BX/RS/SD/XRS Pumps)
UltiMate 3000 Series LPG-3400XRS Pump
UltiMate 3000 Series - Variable Wavelength Detector
UltiMate 3000 Series - Diode Array Detectors
UltiMate 3000 Series - Fluorescence Detectors
UltiMate 3000 Series - ECD-3000RS Electrochemical Detector
Dionex Corona Veo Charged Aerosol Detector