Commands and Tips for Third-Party Devices
Agilent 1100/1200 HPLC System: Pump
In addition to (or different from) the standard pump commands (see Commands for Controlling Dionex Devices Dionex Pumps) and the general commands for the 1100/1200 HPLC systems (see Agilent 1100/1200 HPLC System: General), the following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
Property |
Range/Values |
Description |
0.0 % - 100.0 % |
Changes the solvent composition and determines the gradient profile. |
|
Compressibility |
autom., no compensation, 50..150 |
Compressibility [1e-6/bar] (Available only for the isocratic and quaternary pumps.) |
Compressibility
|
Agilent 1100/1200 pump: autom., no compensation, 50..150 |
Compressibility of the left/right pump [1e-6/bar] (Available only for the binary pumps of the 1100/1200 (SL) system.) |
|
Agilent 1200 binary pump SL: ACN_90_H2O_10 ACN_95_H2O_5 IPA, MeOH, ACN, H2O, and UserDefined |
|
EmptySolventError |
Off / On |
* |
0.000 - 10.000 (quaternary) |
Flow in [ml/min] |
|
Pressure |
Agilent 1100/1200 pump: 0.0 - 400.0 (Flow < 5 ml/min) 200.0 bar (else) |
Pressure in [bar] If one of the pressure limits is exceeded, Chromeleon turns off the flow. |
Agilent 1200 binary pump SL: 0.0 - 600.0 |
||
|
Value |
Reports the current pump pressure. |
|
LowerLimit |
Lower Pressure Limit. Note: The pump flow will not be stopped, if the current pump pressure never exceeds the defined lower limit. This means that a 'real' pressure drop must occur to trigger a flow stop. |
|
UpperLimit |
Upper Pressure Limit |
PrimaryChannel |
autom., A, B, C, D |
(Quaternary LPG pump only) Select the channel from which the pump shall draw liquid first. The pump starts with drawing half the volume required for the composition, then draws the required volume from the other channels, and then draws the second half of the volume required for the first channel. Automatic selects the channel with the maximum percentage. |
SolventFillingX |
0.00 l - 1000.00 l |
* Enter the volume of solvent originally added to the bottle. (Depending on the pump type, X = A1, A2, B1, B2 or A, B, C, D, respectively.) |
SolventLeftPump |
A1, A2 |
Eluent of the left pump. (Available only for the binary pump.) |
SolventNotReady |
0.05 l - 1000.00 l |
* Determine the minimum liquid level. |
SolventRightPump |
B1, B2 |
Eluent of the right pump. (Available only for the binary pump.) |
Stroke |
autom., 20..100 µl |
Pump stroke (Available only for the isocratic and quaternary pumps.) |
StrokeLeftPump |
autom., 20..100 µl |
Stroke of the left pump. (Available only for the binary pump.) |
StrokeRightPump |
autom., 20..100 µl |
Stroke of the right pump. (Available only for the binary pump.) |
* For more information about these commands, refer to Agilent 1100/1200 HPLC System: Checking the Solvent Liquid Level.
Rear Seal Wash Commands
The following commands are available only if the pump is equipped with the rear seal wash option and the Rear Seal Wash check box is selected on the Options tab page in the Server Configuration:
Tip:
This function is available only if the Rear Seal Wash option is installed for the pump.
Command |
Description |
System |
Specifies a wash option: Off, Single, or Interval. |
Duration |
Specifies the duration of a single wash. This command is associated with the Single function of the System command. |
Interval |
Specifies the duration of the entire wash interval, i.e., the idle period and run time of the pump. This command is associated with the Interval function of the System command. |
On Time |
Specifies the actual run time of the pump during a wash period. This command is associated with the Interval function of the System command. Tip: The time specified here may not exceed 70% of the duration specified under Interval. |
Diagnostic Commands
Note:
Pump diagnostic commands are available for Agilent 1100 HPLC Systems only.
The Administrator Help section provides installation instructions for the 1100 and 1200 HPLC systems; refer to Hardware Installation:
Agilent 1100 HPLC System: Installation
Agilent 1200 HPLC System: Installation