Dionex Detectors
Dionex PDA/PDA-100/PDA-3000 Photodiode Array Detectors
In addition to the standard commands available for all detectors (see Dionex Detectors), the PDA, PDA-100, and PDA-3000 detectors provide several additional control commands (please note that the display Filter level determines which commands are displayed).
UV_VIS Channel Commands
Command |
Description |
Sets the bandwidth (the wavelength range at which the chromatogram is measured). |
|
Sets the offset for the analog output signal. |
|
Sets the range of a full-scale recorder response (analog output). |
|
Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect. |
|
Sets the reference wavelength. |
3DFIELD Channel Commands
Command |
Description |
Determines how many wavelengths are averaged when collecting 3D data. Bunching (averaging) the signals of adjacent wavelengths will reduce the amount of 3D data stored. |
|
MaxWavelength |
Sets the maximum wavelength of the 3D field. |
MinWavelength |
Sets the minimum wavelength of the 3D field. |
RefBandwidth |
Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect. |
RefWavelength |
Sets the reference bandwidth. |
Other Special Detector Commands
Command |
Description |
AmbientTemp |
Reports the temperature outside the optical assembly. |
AZ_Done |
Reports the status of an autozero command (Busy or Done). |
Sets the rate at which Chromeleon collects digital data points from the detector. |
|
Sets the amount of increased light transmission from autozero (negative absorption) to which the detector responds. This permits peaks to be seen even when the eluent decreases absorbance. |
|
OpticalBenchTemp |
Reports the temperature of the optical bench. |
Sets how quickly the detector responds to a change in signal. |
|
Sets the TTL input signal mode. |
|
Turns the deuterium lamp on and off. |
|
UVLampPowerOnSetting |
Selects whether the deuterium lamp is turned on when the detector power is turned on. The default setting is On. |
Turns the tungsten lamp on and off. |
|
VisLampPowerOnSetting |
Selects whether the tungsten lamp is turned on when the detector power is turned on. The default setting is On. |
UV Calibration Commands
Command |
Description |
DownloadCalibration |
Downloads the selected set of calibration values to the detector. |
DownloadCalibration Parameter |
Selects the set of calibration values (Current, Previous, or Factory) to be downloaded. |
FilterWheel |
Calibrates the filter wheel (PDA-100 (DX-LAN) only). |
Intensity |
Calibrates the intensity of the deuterium lamp. |
LeakDet |
Performs the leak detector calibration (PDA-100 (DX-LAN) only). |
WavelengthCal |
Performs a wavelength calibration. |
WavelenthCalDate |
Reports the date of the wavelength calibration. |
WavelenthCalResult |
Reports the result (pass or fail) of the wavelength calibration. |
UV Diagnostic Commands
Command |
Description |
LeakDetTest |
Performs a test of the leak detector (PDA-100 (DX-LAN) only). |
LeakDetTestDate |
Reports the date of the leak detector test. |
LeakDetTestResult |
Reports the result (pass or fail) of the leak detector test. |
WavelengthVerTest |
Performs wavelength verification via the Holmium Oxide Filter. |
WavelengthVerTestDate |
Reports the date of the wavelength verification. |
WavelengthVerTestResult |
Reports the result (pass or fail) of the wavelength verification. |
System Wellness and Predictive Performance commands and properties
Tip:
The Chromeleon version and the detector type determine which commands and properties are available for checking the detector wellness and performance.
Command/Property |
Description |
Reports the deuterium lamp's intensity at 254 nm in counts/second. |
|
ModuleAge |
Reports (or resets) the number of hours the detector has been on. |
Qualification |
Provides information about detector qualification. |
*.LastDate |
Reports when qualification was last performed (read-only). |
*.LastOperator |
Reports the name of the user who performed the last qualification (read-only). |
*.Interval |
Specifies the period of time (in days) between qualifications. |
*.WarningPeriod |
Specifies the number of days before qualification expiration that Chromeleon will display a reminder that requalification will be necessary. |
*.GracePeriod |
Indicates for how many more days the user can continue to operate the detector after failing to perform a scheduled qualification. |
QualificationDone |
Selecting this command after qualification sets *.LastDate to the current date and sets *.LastOperator to the name of the currently logged in user. |
UVLampAge |
Reports (or resets) the number of hours the deuterium lamp has been on. |
VisLampAge |
Reports (or resets) the number of hours the tungsten lamp has been on. |
VisLampIntensity |
Reports the tungsten lamp's intensity at 520 nm. Reported in counts per second. |
The Administrator Help section provides information about how to install the PDA, PDA-100, or PDA-3000 detector; refer to Hardware Installation:
ICS-Series Photodiode Array Detector: Overview
PDA-100 (DX-LAN) Detector: Overview
Photodiode Array Detector (PDA-100 (USB)/PDA-3000): Installation.