Dionex Detectors
Dionex PDA/PDA-100/PDA-3000 Photodiode Array Detectors

In addition to the standard commands available for all detectors (see  Dionex Detectors), the PDA, PDA-100, and PDA-3000 detectors provide several additional control commands (please note that the display Filter level determines which commands are displayed).

UV_VIS Channel Commands

 

Command

Description

Bandwidth

Sets the bandwidth (the wavelength range at which the chromatogram is measured).

Offset_Level

Sets the offset for the analog output signal.

Recorder_Range

Sets the range of a full-scale recorder response (analog output).

RefBandwidth

Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect.

RefWavelength

Sets the reference wavelength.

 

 

3DFIELD Channel Commands

 

Command

Description

BunchWidth

Determines how many wavelengths are averaged when collecting 3D data. Bunching (averaging) the signals of adjacent wavelengths will reduce the amount of 3D data stored.

MaxWavelength

Sets the maximum wavelength of the 3D field.

MinWavelength

Sets the minimum wavelength of the 3D field.

RefBandwidth

Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect.

RefWavelength

Sets the reference bandwidth.

 

Other Special Detector Commands

 

Command

Description

AmbientTemp

Reports the temperature outside the optical assembly.

AZ_Done

Reports the status of an autozero command (Busy or Done).

Data_Collection_Rate

Sets the rate at which Chromeleon collects digital data points from the detector.

NegativeAbsorptionLevel

Sets the amount of increased light transmission from autozero (negative absorption) to which the detector responds. This permits peaks to be seen even when the eluent decreases absorbance.

OpticalBenchTemp

Reports the temperature of the optical bench.

Rise_Time

Sets how quickly the detector responds to a change in signal.

TTL_Input_Mode

Sets the TTL input signal mode.

UV_Lamp

Turns the deuterium lamp on and off.

UVLampPowerOnSetting

Selects whether the deuterium lamp is turned on when the detector power is turned on. The default setting is On.

Visible_Lamp

Turns the tungsten lamp on and off.

VisLampPowerOnSetting

Selects whether the tungsten lamp is turned on when the detector power is turned on. The default setting is On.

 

UV Calibration Commands

 

Command

Description

DownloadCalibration

Downloads the selected set of calibration values to the detector.

DownloadCalibration Parameter

Selects the set of calibration values (Current, Previous, or Factory) to be downloaded.

FilterWheel

Calibrates the filter wheel (PDA-100 (DX-LAN) only).

Intensity

Calibrates the intensity of the deuterium lamp.

LeakDet

Performs the leak detector calibration (PDA-100 (DX-LAN) only).

WavelengthCal

Performs a wavelength calibration.

WavelenthCalDate

Reports the date of the wavelength calibration.

WavelenthCalResult

Reports the result (pass or fail) of the wavelength calibration.

 

UV Diagnostic Commands

 

Command

Description

LeakDetTest

Performs a test of the leak detector (PDA-100 (DX-LAN) only).

LeakDetTestDate

Reports the date of the leak detector test.

LeakDetTestResult

Reports the result (pass or fail) of the leak detector test.

WavelengthVerTest

Performs wavelength verification via the Holmium Oxide Filter.

WavelengthVerTestDate

Reports the date of the wavelength verification.

WavelengthVerTestResult

Reports the result (pass or fail) of the wavelength verification.

 

System Wellness and Predictive Performance commands and properties

 Tip:  

The Chromeleon version and the detector type determine which commands and properties are available for checking the detector wellness and performance.

 

Command/Property

Description

LampIntensity

Reports the deuterium lamp's intensity at 254 nm in counts/second.

ModuleAge

Reports (or resets) the number of hours the detector has been on.

Qualification

Provides information about detector qualification.

*.LastDate

Reports when qualification was last performed (read-only).

*.LastOperator

Reports the name of the user who performed the last qualification (read-only).

*.Interval

Specifies the period of time (in days) between qualifications.

*.WarningPeriod

Specifies the number of days before qualification expiration that Chromeleon will display a reminder that requalification will be necessary.

*.GracePeriod

Indicates for how many more days the user can continue to operate the detector after failing to perform a scheduled qualification.

QualificationDone

Selecting this command after qualification sets *.LastDate to the current date and sets *.LastOperator to the name of the currently logged in user.

UVLampAge

Reports (or resets) the number of hours the deuterium lamp has been on.

VisLampAge

Reports (or resets) the number of hours the tungsten lamp has been on.

VisLampIntensity

Reports the tungsten lamp's intensity at 520 nm. Reported in counts per second.

 

The Administrator Help section provides information about how to install the PDA, PDA-100, or PDA-3000 detector; refer to Hardware Installation:

 ICS-Series Photodiode Array Detector: Overview

 PDA-100 (DX-LAN) Detector: Overview

 Photodiode Array Detector (PDA-100 (USB)/PDA-3000): Installation.