Dionex Pumps
PP-150 Preparative HPLC Pump

In addition to the standard pump commands (see Commands for Controlling Dionex Devices  Dionex Pumps), the commands and properties from the table below are available (please note that the display Filter level determines which commands and properties are displayed).

 

Property

Values

Description

%B, %C, %D

0.0 - 100.0 %

Use %B, %C, and/or %D to determine the solvent composition.

Flow

0.000 - 9.999 ml/min

Sets the flow rate.

Motor

On/Off

(Only available for the first pump) Turns the (all) pump(s) on or off

ModelVariant

Unknown
Isocratic
LPG
HPG

Indicates the pump's operating mode (read-only). Set the operating mode on the Pumps tab page in the Server Configuration program.

Pressure

0 - 432 bar

Indicates the pressure (read-only).

Pressure.LowerLimit

0 - 392 bar

Sets the lower pressure limit.

Pressure.UpperLimit

10 - 432 bar

Sets the upper pressure limit.

Solvent
(Pump: PASV, PBSV, PCSV)

A, B, C, D

(For pumps with connected FCV Solvent Selector) Selects the solvent.

 

 Tip:

To use the Motor On/Off, CycleMode, and Solvent properties, you have to enter them manually in the PGM Editor.

 

Observe the following when creating a program:

When the pressure unit is psi, always add the entry in brackets, e.g.

 PrepPump.Pressure.LowerLimit = 400 [psi]

Try to avoid Delay. Instead, use the retention time whenever possible.

In triggers, use AbortSample instead of End.

Use Protocol instead of Message.

For information about how to install the pump, refer to Hardware Installation  PP-150 Preparative HPLC-Pump: Overview.