Installing and Controlling Third-Party Devices
Agilent 7673 and 7683 Autosamplers: Dual Inject
To double the sample throughput, the 7673 or 7683 GC autosamplers support dual injection, allowing you to inject two samples simultaneously from different sample positions. Make sure that the following conditions are met:
Hardware Installation
Devices: |
|
Columns: |
Two |
Injectors: |
Front and Rear Injector |
Detectors: |
Detector A and Detector B Note: Usually, identical columns, injectors, and detectors are used. |
Device |
In addition to the standard cabling for the HP 5890 (see HP 5890: Installation) and the 7673 autosampler (see Agilent 7673 Autosampler on the HP 5890 GC Using the HP G1512 Controller-Box or Agilent 7673 Autosampler at HP 5890 Using the HP 18594 Controller Box), the injection signal must be sent to the UCI. |
Settings in the Server Configuration
For information about the program, refer to Software Installation and Communication The Server Configuration Program.
Timebases: |
For data acquisition with two detectors, two timebases are required: one controlled timebase, for example, Timebase 1, and one non-controlled timebase, for example, Timebase 2. To create the timebases, install the following Device Drivers: |
|
Timebase 1: |
Specify which options are installed. For installation information, refer to: Agilent 7673 Autosampler on the HP 5890 GC Using the HP G1512 Controller Box Agilent 7673 Autosampler on the HP 5890 GC Using the HP 18594 Controller Box. |
|
Timebase 2: |
Open the Signal Configuration dialog box by double-clicking the corresponding signal on the Signals tab page. Assign the signal the channel of UCI Universal Chromatography Interface to which the analog cable of the second detector is connected. Name the signal in the Signal Name field according to the selected detector type; for example, FID_2. Under Inject Port on the General tab page, select the input of the UCI to which the injection signal is assigned. Enter a name; for example, HP7673_Inject. |
|
Signal |
Timebase 1 (HP 5890) Signal GC_1 « UCI100_CHANNEL1 « Signal GC_2 « UCI100_CHANNEL2 « (Required for detector control) Timebase 2 (Integrator Driver) Signal INT_1 « UCI100_CHANNEL3 « |
Device - HP 5890 Signal from detector A
Signal from detector B |
For information about the Dual Inject option, refer to Practical Tips for Device Control Injecting Two GC Samples Simultaneously (7673 and 7683 Autosamplers) in the User Help section.