Sampler
Enter the configuration of the installed autosampler:
Option |
Description |
|
|
|
|
Installed |
Make sure that the Installed check box is selected if a Shimadzu autosampler is configured with the system. |
|
|
|
|
Device Name: |
Indicates the name under which the instrument is identified in the installation environment (default: Sampler; also, refer to Naming Devices...). To control the instrument via the existing Control Panels, accept the default name. If you do not accept the default name, you may have to edit the PGM Files and re-link the corresponding control of the control panel. (Also, refer to How to: Controlling Devices from the Control Panel Linking a Control to a Device in the User Help section.) |
|
|
|
|
Tray Type |
Select the installed rack type: |
|
175 Vials (1 mL) |
105 Vials (1.5 mL) |
|
70 Vials (1.5 mL) |
50 Vials (4 mL) |
|
96 Well Plates |
384 Well Plates |
|
|
384 Deep Well Plates |
|
|
Changer Rack (only available if a rack changer is connected with the autosampler) |
|
96 Deep Well Plates |
|
|
|
|
|
SIL-10ADvp |
150 Vials |
One 96 Well Plate |
70 Vials |
Two 96 Well Plates |
|
60 Vials |
One 384 Well Plate |
|
100
Vials + R101-103 |
Two 384 Well Plates |
|
70 Vials + R101-103 |
|
|
|
|
|
SIL-10A, SIL-10AF SIL-10Ai, SIL-10AXL
|
100 Vials |
25 Vials |
80 Vials |
One 96 Well Plate |
|
60 Vials |
Two 96 Well Plates |
|
50 Vials |
|
|
|
|
|
Standard
100 無 Loop |
Select this option if a standard sample loop is installed (100 無; 50 無 for the SIL-10ADvp). This setting is the default. |
|
Optional
2000 無 Loop |
Select this option if a 2000 無 sample loop (or 500 無 sample loop for the SIL-10ADvp) is installed. Attention: When changing the volume of the sample loop in Chromeleon, make sure to also adjust the value on the front panel of the instrument. It is vital that the same value is used both on the instrument and in Chromeleon when configuring the loop volume for the first time or editing an existing configuration. |
|
|
|
|
No
Prep Loop |
Make sure that this option is selected if you have no sample preparation loop installed. This setting is the default. If this option is selected, no sample preparation options are provided. |
|
Prep
Loop |
Select this option if you have a sample preparation loop installed. |
|
|
|
|
Cooling Option |
Select this option if you are using an autosampler that allows sample cooling. (Note: this option is not supported for the SIL-10AXL.) |
|
|
|
|
Loop Injection Kit (only SIL-30AC) |
Select this check box, if the autosampler is equipped with an optional loop injection kit. |
|
|
|
|
Ext. Rinse pump (only SIL-20A, SIL-20AC, SIL-20AXR, SIL-20ACXR) |
Select this check box, if the autosampler is equipped with an optional pump for rinsing the needle and flow lines. |
|
|
|
|
500 無 Syringe 2500 無 Syringe 5000 無 Syringe (only SIL-10A/AF/Ai) |
Select the installed syringe type. The following requirements must be met in order to use the 5000 無 syringe type:
|
Also, refer to Hardware Installation:
Shimadzu LC-10A/LC-20A/LC-30A HPLC System: Installation