Sampler

Enter the configuration of the installed autosampler:

 

Option

Description

 

 

Installed

Make sure that the Installed check box is selected if a Shimadzu autosampler is configured with the system.

 

 

Device Name:

Indicates the name under which the instrument is identified in the installation environment (default: Sampler; also, refer to Naming Devices...). To control the instrument via the existing Control Panels, accept the default name. If you do not accept the default name, you may have to edit the PGM Files and re-link the corresponding control of the control panel. (Also, refer to How to: Controlling Devices from the Control Panel  Linking a Control to a Device in the User Help section.)

 

 

Tray Type

Select the installed rack type:

175 Vials (1 mL)

105 Vials (1.5 mL)

70 Vials (1.5 mL)

50 Vials (4 mL)

96 Well Plates

384 Well Plates

 

384 Deep Well Plates

 

Changer Rack

(only available if a rack changer is connected with the autosampler)

96 Deep Well Plates

 

 

 

SIL-10ADvp

150 Vials

One 96 Well Plate

70 Vials

Two 96 Well Plates

60 Vials

One 384 Well Plate

100 Vials + R101-103
(100 vials + 3 reagent vials)

Two 384 Well Plates

70 Vials + R101-103

 

 

 

SIL-10A, SIL-10AF

SIL-10Ai, SIL-10AXL

 

100 Vials

25 Vials

80 Vials

One 96 Well Plate

60 Vials

Two 96 Well Plates

50 Vials

 

 

 

Standard 100 無 Loop
Standard 50 無 Loop

Select this option if a standard sample loop is installed (100 無; 50 無 for the SIL-10ADvp). This setting is the default.

Optional 2000 無 Loop
Optional 500 無 Loop

Select this option if a 2000 無 sample loop (or 500 無 sample loop for the SIL-10ADvp) is installed.

 Attention:

When changing the volume of the sample loop in Chromeleon, make sure to also adjust the value on the front panel of the instrument.

It is vital that the same value is used both on the instrument and in Chromeleon when configuring the loop volume for the first time or editing an existing configuration.

 

 

No Prep Loop
(only SIL-10ADvp)

Make sure that this option is selected if you have no sample preparation loop installed. This setting is the default. If this option is selected, no sample preparation options are provided.

Prep Loop
(only SIL-10ADvp)

Select this option if you have a sample preparation loop installed.

 

 

Cooling Option

Select this option if you are using an autosampler that allows sample cooling. (Note: this option is not supported for the SIL-10AXL.)

 

 

Loop Injection Kit

(only SIL-30AC)

Select this check box, if the autosampler is equipped with an optional loop injection kit.

 

 

Ext. Rinse pump

(only SIL-20A, SIL-20AC, SIL-20AXR, SIL-20ACXR)

Select this check box, if the autosampler is equipped with an optional pump for rinsing the needle and flow lines.

 

 

500 無 Syringe

2500 無 Syringe

5000 無 Syringe

(only SIL-10A/AF/Ai)

Select the installed syringe type.

The following requirements must be met in order to use the 5000 無 syringe type:

  • firmware version 4.10 or later

  • CPU board version 3.51

 

 

Also, refer to Hardware Installation:

 Shimadzu LC-10A/LC-20A/LC-30A  HPLC System: Installation

 Shimadzu LC-10A/LC-20A and LC-2010 HPLC Systems: Racks

 Shimadzu LC-2010 HPLC System: Installation