Detector Control
Modifying the Signal Parameters of a UV Channel
Modify one or several signal parameters to improve the signals provided by the UV Detector:
Signal |
Description |
Comment |
Averaging. |
Use this parameter to improve the signal-to-noise ratio. |
|
Changes the bandwidth with which the signal is recorded. |
If necessary, increase the bandwidth to improve the Signal to Noise Ratio. |
|
Delta |
Indicates the difference between the current signal value and the signal value one second before. |
The value cannot be changed. |
MaxAutoStep |
Sets the maximum step width for the auto setting. |
The highest possible value is the default: 5.1 s. |
Polarity |
Changes the polarity of the output signal. |
For some detectors, you can use this parameter to change the polarity of the output signal. |
Retention |
Indicates the current retention time |
The value cannot be changed. |
Signal |
Value indicates the current value of a Signal Use UpperLimit and LowerLimit to set the signal acquisition limits. |
If signal acquisition limits are not met, the running batch is stopped and the Emergency Program is started. |
Changes the step. |
Select the step such that the narrowest peak contains approximately 20 datapoints |
|
Changes the reference bandwidth. |
If necessary, increase the bandwidth to improve the signal-to-noise ratio. |
|
Changes the recorded wavelength. |
Note the possibilities for wavelength switching (see Switching the Wavelength). |