Detector Control
Modifying the Signal Parameters of a UV Channel

Modify one or several signal parameters to improve the signals provided by the UV Detector:

 

Signal
Parameter

 

Description 

 

Comment 

Average

Averaging.

Use this parameter to improve the signal-to-noise ratio.

Bandwidth

Changes the bandwidth with which the signal is recorded.

If necessary, increase the bandwidth to improve the Signal to Noise Ratio.

Delta

Indicates the difference between the current signal value and the signal value one second before.

The value cannot be changed.

MaxAutoStep

Sets the maximum step width for the auto setting.

The highest possible value is the default: 5.1 s.

Polarity

Changes the polarity of the output signal.

For some detectors, you can use this parameter to change the polarity of the output signal.

Retention

Indicates the current retention time

The value cannot be changed.

Signal

Value indicates the current value of a Signal

Use UpperLimit and LowerLimit to set the signal acquisition limits.

If signal acquisition limits are not met, the running batch is stopped and the Emergency Program is started.

Step

Changes the step.

Select the step such that the narrowest peak contains approximately 20 datapoints

RefBandwidth (Reference Bandwidth)

Changes the reference bandwidth.

If necessary, increase the bandwidth to improve the signal-to-noise ratio.

Wavelength

Changes the recorded wavelength.

Note the possibilities for wavelength switching (see  Switching the Wavelength).