Commands and Tips for Third-Party Devices
Waters 996 and 2996 PDA Detectors

 Tip: 

Before data acquisition can start, the detector needs some time to load the parameter settings. That is why a Wait command is required before data acquisition is started.

In addition to the standard commands, the Waters 996/2996 PDA detector driver supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):

 

Property

Min

Max

Description

ExposureTime

Auto;
11 ms

500 ms

Time during which a single diode is exposed to light. (The value cannot exceed the Step value - see below.)

 Tip: 

Do not change the default ('Auto') unless absolutely required.

Interpolate656nm

Off

On

On: To protect the photodiodes from over-saturation, the signal at 656.1 nm (Balmer line for deuterium) is ignored (visible only on the Expert level).

Lamp

Off

On

Turns the lamp on or off.

 

UV Channel Commands

 

Property

Min

Max

Default

Description

Average

Off

On

On

Allows signal averaging.

Bandwidth

1.2 nm

30 nm

1.2 nm

 

Step

Auto, 0.01 s

4.80 s

Auto

Indicates the interval between two data points. (The value must exceed the ExposureTime value.)

Wavelength

190 nm

800 nm

240.0 nm

Indicates the wavelength at which the detector measures the signal.

 

 Tip:  

After a Reset command, it takes 90 s until the detector is ready to execute further commands.

 

Commands and Properties of the 3D Field

 

Property

Min

Max

Default

Description

BunchWidth

1.2 nm

60 nm

1.2 nm

Describes the bandwidth of a bunch of diodes.

MaxWavelength

MinWavelength

799.6 nm

799.6 nm

Indicates the upper limit of the measuring range. (The setting cannot be changed during data acquisition.)

MinWavelength

190.0 nm

MaxWavelength

190.0 nm

Indicates the lower limit of the measuring range. (The setting cannot be changed during data acquisition.)

Step

0.1 s
0.2 s
0.5 s
1.0 s
2.0 s
3.0 s
4.0 s

0.1 s

Indicates the interval between two data points. (The value must exceed the ExposureTime value.)

 

 Note:  

Due to the minimum Bunch Width of 1.2 nm, the minimum and/or maximum wavelength can only have the following value: 190.0 nm + n*1.2 nm (where n = integer). If you enter a different value, the next lower value is used for the minimum wavelength and the next higher value is used for the maximum wavelength. This is logged in the Audit Trail.

Calibration Commands and Properties

The following commands and properties are provided for calibrating and validating the wavelength of the 996 and 2996 PDA detectors.

In addition, a dedicated sub-panel is provided in Chromeleon which can be used for validation and calibration of the wavelength. The sub-panel can be accessed in the Waters 996 / 2996 PDA section using the Calibration button provided on the Waters2690_996PDA.pan or Waters2790_996PDA.pan which are available in the Chromeleon Browser under Panels/LC_Other.

 

Property

Description

CalibrationMaxWl

Specifies the center wavelength of the first diode in the array.

On the calibration sub-panel of the 996/2996 PDA, the center wavelength of the first diode can be entered in the Array start box.

CalibrationMinWl

Specifies the center wavelength of the second diode in the array.

On the calibration sub-panel of the 996/2996 PDA, the center wavelength of the second diode can be entered manually in the Array end box.

WavelengthCal

Starts calibration of the wavelength.

WavelengthVal

Starts validation of the wavelength.

WlDeviation_A

Reports the deviation of the first (D-alpha) calibration peak.

WlDeviation_B

Reports the deviation of the second (D-beta) calibration peak.

 

The standard relay commands are available to control the relays of the Waters 996 and 2996 PDA detectors (refer to Practical Tips for Device Control  Relay, TTL, and Remote Input Commands).

 

For information about how to install the Waters 996 and 2996 PDA detectors, refer to Hardware Installation  Waters 996 and 2996 PDA Detectors: Installation.