Commands and Tips for Thermo Scientific
Devices
ThermoQuest TRACE GC (2000/Ultra) and Focus GC Commands
A large number of commands is available for the individual options of the TRACE and FOCUS GCs. Only the most important of them are described below (please note that the display Filter level determines which commands and properties are displayed). They are common to the respective device types (for example, the different detectors). For a short description of all GC commands, refer to the corresponding Commands view (press the F8 key in the program) or the Properties/Link box.
GC
Command |
Min |
Max |
Default |
Description |
AutoPrepRun |
Off |
On |
Off |
Allows the GC to automatically prepare the run. |
AutoStart |
Off |
On |
Off |
Allows the GC to automatically start the run when it becomes ready. Note: Before enabling AutoStart, make sure to do the following:
|
Cryo |
Off |
On |
Off |
ON: The cryo valve will operate when needed. OFF: The cryo valve is disabled. (The property is visible only if the cryo option is enabled.) |
Range |
0 |
3 |
1 |
Sets the output range of the detector signal. Note: Logarithmic scale with 10^ nA. |
RawSignalValue |
|
|
|
Value of the detector signal as shown on the keypad. |
SignalSource |
|
|
|
Sets the signal source of the relevant GC channel. The following settings are available: None (no assignment) Tip: If all
three available detectors are installed, assign the relevant signal
sources as follows: |
TempCtrl |
OFF |
ON |
ON |
Turns the oven's temperature control on or off. |
Temperature |
|
|
|
The allowed maximum temperature change is 120°C/min (1200°C/min for the TRACE UltraFast). A maximum of 6 ramps per program are possible. |
.Nominal |
30 |
Note 2 |
50 |
Nominal oven temperature in [°C]. |
.Value |
30 |
Note 2 |
N/A |
Actual oven temperature in [°C] (read-only). |
.UpperLimit |
70 |
Note 2 |
350 (FOCUS) |
Upper limit for the oven temperature in [°C]. |
.LowerLimit |
30 |
50°C |
30 |
Lower limit for the oven temperature in [°C]. |
Note 1: TRACE GC: Depending on the cryo settings, the minimum temperature can be 30°C, -55°C, or -99°C.
Note 2: TRACE GC: Depending on the cryo settings,
the maximum temperature can be 450°C or 500°C.
FOCUS GC: Depending on the cryo settings, the maximum temperature can be
350°C.
Inlet
Command |
Range |
Default |
Description |
TempCtrl |
ON, OFF |
ON |
Turns temperature control on or off. |
Temperature |
|
|
|
.Nominal |
30–450
(TRACE) |
50
(TRACE) |
Nominal inlet temperature in [°C] |
.Value |
30–450
(TRACE) |
N/A |
Actual inlet temperature in [°C] (read-only). |
Mode |
N/A |
Split |
Inlet mode: Split or Splitless |
Detector(s)
Command |
Range |
Default |
Description |
|
|
Sets the signal output to 0. |
|
TempCtrl |
ON, OFF |
ON |
Turns temperature control on or off. |
Temp |
|
|
|
.Nominal |
30–450
(TRACE) |
150 |
Nominal detector temperature in [°C] |
.Value |
30–450
(TRACE) |
N/A |
Actual detector temperature in [°C] (read-only). |
Column(s)
Command |
Range |
Default |
Description |
PressureCtrl |
ON, OFF |
ON |
Turns the control of the column gas pressure on or off. |
Pressure |
|
|
|
.Nominal |
0.03– 10.00 |
N/A |
Nominal pressure in the column [bar]. |
.Value |
0.03–10.00 |
N/A |
Actual pressure in the column [bar] (read-only). |
FlowCtrl |
ON, OFF |
ON |
Turns the column flow control on or off. |
Flow |
|
|
|
.Nominal |
0.0–50.0
(TRACE) |
N/A |
Nominal flow in the column [ml/min]. |
.Value |
0.0–50.0
(TRACE) |
N/A |
Actual flow in the column [ml/min]. |
Valves and External Events
The following commands and properties are available if a valve or external event is configured on the Valves dialog page in the Chromeleon Server Configuration.
Command/Property |
Range |
Description |
Default (FOCUS GC only) |
Off (0) or On (1) |
(External events only) Specifies the default value of the external event (the setting is valid for as long as the FOCUS GC is not performing a run). Note: The following interaction exists between the Default and State command: When a Default value is specified, State is automatically set to Off. If you want to specify a Default value and State = On at the same time, you must reset State to On. |
On/Off |
|
Sets the valve or external event state either to On or Off. Duration: (Optional) Specifies a time limit after which the valve or event state is reset to the initial value (i.e., either On or Off). Range: 0.6–60000.0 s. |
State |
Off (0) or On (1) |
(External events only) Sets the event state. |
Type |
ON |
Displays the valve type (GSV or External) as specified in the Chromeleon Server Configuration (read-only). |
Program Tips:
Valve and external event commands cannot be executed manually; they can only be used in a sample program or a program in which a GC run is performed.
Valve and external event commands cannot be set at a negative program time (t < 0).
All commands and parameters in a sample PGM must be set at program time t > 0.
If a Default value is specified, a State value should be set at the same time step (FOCUS GC only).
Dual Injection
The following commands are available if the GC is configured for dual operation:
Further Information
The Administrator Help section provides more information about the TRACE and FOCUS GCs. Refer to Hardware Installation: