Commands and Tips for Thermo Scientific Devices
Thermo Scientific TriPlus RSH / 100 LS Autosampler: Tips for Sequences

 Note:

Headspace, SPME and ITEX mode are only available for the TriPlus RSH.

Restrictions for Headspace, SPME and GC Liquid sequences

When running an analysis using a TriPlus RSH sampler, the following condition must be met:

Sample Overlap (HS, SPME)

Overlapping sample preparation (see Autosampler Control Overlapping Samples) is supported for headspace and SPME (Solid Phase Microextraction) injections and can be viewed on the Control Panel of the TriPlus RSH Autosampler.

For optimal use of the sampler's overlapping sample preparation, Chromeleon sequences should meet the following conditions:   

  1. Each run can contain only one program. If you want to use a different program, you must set up a new run.

  2. Use consecutive injections with a different position number for each sample vial. Overlapping samples is not possible for multiple injections taken from the same sample vial.

  3. Use the same injection volume for all injections. It can be set in the sequence or in the Program Wizard (see The Control Program The Program Wizard) of the TriPlus RSH (click "Launch Editor" in the wizard) if the "Extract injection volume from TriPlus RSH method" option is enabled in the Server Configuration program.

 Note:

Successful overlapping also depends on a number of other factors, for example:

The Chromeleon sequence is split into several runs if:

Injection Volumes

When specifying injection volumes in the sequence, use the following recommended settings to prevent injection volume rounding:

 Note:

If an injection volume specified in the sequence does not comply with the above recommended settings, it will be rounded to the nearest allowed value and a warning containing the details will be written to the audit trail.

 

For installation details, refer to Hardware Installation:

 Thermo Scientific TriPlus RSH Autosampler