Shimadzu LC-10A, LC-20A, LC-30A and LC-2010 HPLC Systems: Troubleshooting

LC-10A, LC-20A and LC-30A

If the manual inject valve stays in the ‘Inject’ position at the end of the current Sample / start of next sample in a sequence, the next run is immediately started by the Shimadzu hardware (that is: without waiting for the valve to be switched to ‘Load’  and back to ‘Inject’ position.

As a result, It is not possible to let the system run into the next sample automatically and then wait until the user switches the valve into Bypass state and back to Inject.

LC-10A

LC-10A and LC2010

SIL-30A and SIL-30AC Autosamplers

When stopping a batch abruptly during sample pretreatment, in some rare cases the sampler may not perform any wash operations. After resuming the sequence the next sample is started (indicated by highlighting the sample in the sample table), however, the sample will get stuck forever.

LC-2010

SPD-M20A

For a list of error messages that might appear in Chromeleon refer to the Error Levels tab page in the Server Configuration program.

 

For information about how to install the Shimadzu LC-10A HPLC System, refer to Hardware Installation  Shimadzu LC-10A/LC-20A HPLC System: Installation in the Administrator Help section.