Dionex Nano Cap Systems
Dionex UltiMate 3000 Series NCS-3500RS: Column Compartment
The following commands and properties are provided for the column compartment of the NCS-3500RS (please note that the display Filter level determines which commands and properties are displayed):
Commands and properties of the main device ("ColumnOven"):
Command/Property |
Description |
|
Brightness |
Adapts the screen brightness of the lower device display to your requirements. |
|
Contrast |
Adapts the screen contrast of the lower device display to your requirements. |
|
Door |
Reports whether the compartment door is open or closed. |
|
DriverID |
Unambiguous driver name. |
|
EquilibrationTime |
The device enters the Ready state only if the deviation between the current temperature and the set point temperature does not exceed the maximum deviation specified under ReadyTempDelta for the time specified here (see also ReadyTempDelta and Ready). |
|
GasLeak |
Reports whether the gas sensor has detected a leak. |
|
GasLeakSensor |
Sets the sensitivity with which the sensor responds to gas. Click the arrow in the input box and select an option from the list. Possible settings are: Low, Standard, High: The Status LED on the front panel is red and a message appears in the Audit Trail. If LeakAlarmMode was set to Audible, a beep sounds additionally. Off: Deactivates gas leak detection. |
|
GasLeakSensorState |
Reports whether the sensor is ready for leak detection (On, Off, Heatup, SwitchOn, DoorOpen, DoorClosed). |
|
HumidityLeak |
Indicates whether the humidity sensor has detected a leak. |
|
HumidityLeakSensor |
Sets the sensitivity with which the sensor responds to humidity. Click the arrow in the input box and select an option from the list. Possible settings are: Low, Standard, High: The Status LED on the front panel is red and a message appears in the Audit Trail. If LeakAlarmMode was set to Audible, a beep sounds additionally. Off: Deactivates humidity leak detection. |
|
HumidityLeakSensorState
|
Reports whether the sensor is ready for leak detection (On, Off, Heatup, DoorOpen, DoorClosed). |
|
Leak |
Indicates whether at least one leak sensor has reported a leak (see also HumidityLeak and GasLeak). |
|
LeakAlarmMode |
Specifies whether the device issues an acoustic alert in case of a leak. Audible: A beep sounds in case of a leak. Silent: The Status LED on the front panel is red, a message appears in the Audit Trail, but no beep sounds. |
|
LeakSensorState |
Reports whether both sensors are ready for leak detection (Ready, NotReady). |
|
Ready |
Reports whether the device is ready to operate (Ready or NotReady; see also ReadyTempDelta, EquilibrationTime). |
|
ReadyTempDelta |
If the current temperature deviates from the temperature set point by more than the ReadyTempDelta, the device enters the NotReady state. If the ReadyTempDelta is set to None, Chromeleon does not check whether the set point temperature deviates from the actual temperature (see also Ready and EquilibrationTime). |
|
Temperature
|
Reports the current temperature of the column oven. The following commands are available if in the command tree: |
|
Value |
Current signal value (read-only) |
|
Nominal |
Nominal temperature |
|
UpperLimit, LowerLimit |
If the current signal value is outside these limits, a warning appears in the Audit Trail. |
|
TempCtrl |
Turns temperature control on or off. |
|
ValveLeft |
Indicates and sets the position of the left or right valve. For example: |
|
Moving |
The valve is switching. |
|
6_1/10_1 |
Indicates the fluid connections in the valve: ports 6 (10) and 1 are connected. |
|
1_2 |
Indicates the fluid connections in the valve: ports 1 and 2 are connected. |
|
NotConnected |
The control electronics is not connected; contact Dionex Service. |
|
Error |
An error occurred. Check the device. |
Commands and Properties for the temperature channel ("ColumnOven_Temp")
This channel is available only if the Use Channel check box is selected in the Properties dialog box (Oven tab page) in the Server Configuration program. If the check box is selected, Chromeleon records the temperature as a separate channel. The channel name is the name entered on the Oven tab page in the Channel Name dialog box.
Click the plus sign beside the name to display the items underneath:
Command/Property |
Description |
AcqOn/Off |
Starts/terminates data acquisition. |
Average |
Averages all measured values over the step interval. Default: On. Off records only the last point of each interval. |
AverageValue |
Indicates the average signal value since the last start of data acquisition (AcqOn). |
Delta |
Reports the signal slope, that is, the difference between the current value and the value one second ago. This is useful for triggers. |
MaxAutoStep |
Maximum step rate for Auto Step Mode. |
Retention |
Reports the retention time of the signal (read-only). |
Signal |
Signal value. |
Step |
Sets the step for data acquisition; range: 0.01 - 4.80 s; Auto selects the best step dynamically. |
System Wellness commands and properties (ColumnOven_Wellness)
Tip:
Note that the installed Chromeleon version, the device type and configuration determine which System Wellness commands and properties are available.
Command/Property |
Description |
HeatingPower |
Reports the current power consumption of the heating unit expressed as a percentage of the maximum performance. |
HeatingWorkLoad |
Reports the total heating workload since the column compartment was operated for the first time. |
OperatingHours |
Counts the operating hours since the column compartment was operated for the first time (excluding the operating hours during standby mode). |
PowerOnTime |
Counts the operating hours since the column compartment was powered up (or resumed operation after standby). |
Qualification |
Provides information about system qualification. |
*.LastDate |
Reports the date when qualification was performed last. |
*.LastOperator |
Reports the name of the person who performed the qualification. |
*.Interval |
Specifies the interval (in days) when the next system qualification is due. |
*.WarningPeriod
|
Specifies the period during which a warning appears before the next service is due. |
*.GracePeriod
|
Indicates the period during which the device may be operated after the due date for qualification. |
QualificationDone |
Perform this command after qualification to update the related system data. |
Service |
Provides service information. |
*.LastDate
|
Reports the date of the last service. |
*.Interval |
Specifies the interval (in days) after which a service is due. |
*.WarningPeriod |
Specifies the period during which a warning appears before the next service is due. |
ServiceDone |
After the device has been serviced, select this command to update the related information. |
StandbyHours |
Counts the total operating hours in standby mode since the device was operated for the first time. |
<Left/Right>RotorSealChangeDate |
Reports when the rotor seal was replaced last. |
<Left/Right>RotorSealChanged |
After you have replaced the rotor seal, perform this command to update the related service information. |
<Left/Right>RotorSealSwitches |
Counts how many times the valve has switched since the rotor seal was replaced last. |
*.Value |
Reports the current number of valve movements. |
*.Warning |
A warning appears when the value specified here is reached. If set to "0" no warning appears. |
*.Limit |
An error message appears when the specified value is reached. |
<Left/Right>ValveCatchHead |
This command is helpful during installation of the right or left valve if the valve fails to lock in place. Perform this command to rotate the actuator and thus allow the valve to lock in place at the correct position. Note: This command is only available if a left and/or right valve has been configured on the Oven/Valves tab page in the Server Configuration. |
<Left/Right>ValveSwitches |
Counts how many times the left or right valve switches. |
*.Value |
Reports the current number of valve switches. |
*.Warning |
A warning appears when the value specified here is reached. If set to "0" no warning appears. |
*.Limit |
An error message appears when the specified value is reached. |
<Left/Right>ValveChanged |
After you have replaced the valves, perform this command to update the related service information. |
<Left/Right>ValveChangeDate |
Reports when the valves were replaced last. |
For information on additional commands and properties for controlling the NCS-3500, refer to:
UltiMate 3000 Series NCS-3500RS: Column ID Properties
UltiMate 3000 Series NCS-3500RS: NC Pump
UltiMate 3000 Series NCS-3500RS: Loading Pump
For information about the commands and properties that are available for other pumps in the UltiMate 3000 pump series, refer to: