Dionex Flow Manager and Column Compartments
Dionex NCS-3500RS Column Compartment: Column ID Properties
The column compartment supports the following column ID properties. The column properties appear underneath the column name specified in the Server Configuration program. (Note that the display Filter level determines which properties are displayed and that, depending on the column compartment type, some commands or properties may not be available.)
Primary column properties
These properties are used for column identification. They are entered once by either the column manufacturer or the user and cannot be changed.
Property |
Description |
BatchNumber |
Batch or lot number of the column. |
CustomId |
You can enter any number you want; for example, the column's inventory number. (Note: The number cannot be changed once it is entered.) |
DateOfManufacture |
Manufacturing date of the column. |
Manufacturer |
Column manufacturer. |
ProductId |
Product number of the column. |
SerialNo |
Serial number of the column. |
Secondary column properties
These properties are entered by the user and can be changed later, if necessary:
Property |
Description |
|
ActiveColumn |
Specify whether the column identification information is updated for the currently used column (Yes or No); |
|
CardState |
Reports whether the memory card is present (read-only): |
|
NoChip |
The card is missing. |
|
DoorOpened |
The door is open. (The card is not writable while the door is open.) |
|
UnknownFormat |
The card format is unknown. |
|
BusBlocked |
One of the cards is not inserted correctly. This may prevent another card from being read, even if it is correctly installed. |
|
DateOfExpiration |
Sets the expiration date of the column. If a date has been entered, the current date is compared with the expiration date during injection or at the beginning of batch processing. If the current date exceeds the expiration date or if batch processing starts less than 24 hours before the expiration date, a warning appears once per Batch. |
|
Description |
Column description. |
|
Diameter |
Column diameter. |
|
EraseChipCard |
Select this command to remove all data from a chip card. |
|
FormatChipCard |
Select this command to format a chip card that contains invalid or corrupt data. |
|
Injections_Limit |
Maximum number of injections allowed for the column. |
|
Length |
Column length. |
|
PackingMaterial |
Packing material of the column. |
|
ParticleSize |
Particle size. |
|
pH_LowerLimit |
Upper and lower pH limits. If both limits are set to the same value, this means that no limit exists. Note: These values are for information purposes only. The device driver cannot check the pH value. |
|
Pressure_UpperLimit |
If the column pressure exceeds the upper pressure limit, a warning appears. Note: Accept the default (0.0) if you do not want to have the system check the pressure. |
|
Temp_LowerLimit |
If the column temperature is above the upper temperature limit or below the lower temperature limit, a warning appears. If both limits are set to the same value (default: 25°C each), this means that no limit exists. (These values also apply to the inactive column.) |
|
VoidVolume |
The column's void volume in [µl]. |
System properties
System properties are read-only and cannot be changed by the user. Chromeleon determines and updates the following properties during each injection:
Property |
Description |
DateOfFirstUse |
Date of the first injection |
DateOfLastUse |
Date of the last injection |
Injections_Counter |
Number of injections |
Pressure_Max |
Maximum pressure measured while the column was in use |
Temp_Min |
Minimum temperature measured while the column was in the column compartment |
Temp_Max |
Maximum temperature measured while the column was in the column compartment |
User properties
The user can change the following properties at any time:
Property |
Description |
Application |
Enter any text to describe the application. |
Comment |
Enter any comment text. |
SystemPressure |
Name of the device to be used for system pressure measurement. |
For information about the special commands supported by the NCS-3500RS column compartments, refer to:
UltiMate 3000 Series NCS-3500RS System: Column Compartment
For information about the standard commands supported for the different column compartments, refer to: