Dionex Detectors
Dionex AD20/AD25 Absorbance Detectors
In addition to the standard commands available for all detectors (see Dionex Detectors), Dionex absorbance detectors provide the following commands (please note that the display Filter level determines which commands and properties are displayed):
Special Commands
Command |
Description |
Sets the rate at which Chromeleon collects digital data points from the detector. |
|
Sets how quickly the detector responds to a change in signal. |
|
UV_Cutover |
For AD20 only: Sets the wavelength above which the second order filter is inserted in the light path. Typically, the UV cutover wavelength is 380nm. |
Turns the deuterium lamp to Off, Low, or High (AD20 detector), or Off or On (AD25 detector). |
|
Turns the tungsten lamp to Off, Low, or High (AD20 detector), or Off or On (AD25 detector). |
|
Sets the wavelength. |
Analog Output Commands
Command |
Description |
For AD20 only: Sets the full-scale analog out voltage. |
|
Sends a positive pulse to the detector’s analog output as an event marker. |
|
Sets the offset for the analog output signal. |
|
Switches the analog output polarity (+/-). |
|
For AD25 only: Allows calibration of the recorder’s response to three detector analog output settings (AU, Zero, Full_Scale) |
|
Sets the range of a full-scale recorder response. |
UV Calibration Commands (For AD25 Only)
Command |
Description |
DownloadCalibration |
Downloads the selected set of calibration values to the detector. |
DownloadCalibration Parameter |
Selects the set of calibration values (Current, Previous, or Factory) to be downloaded. |
LeakDet |
Performs the leak detector calibration. |
WavelengthCal |
Performs a wavelength calibration. |
WavelengthCalDate |
Reports the date of the wavelength calibration. |
WavelengthCalResult |
Reports the result (pass or fail) of the wavelength calibration. |
UV Diagnostic Commands (For AD25 Only)
Command |
Description |
LeakDetTest |
Performs a test of the leak detector. |
LeakDetTestDate |
Reports the date of the leak detector test. |
LeakDetTestResult |
Reports the result (pass or fail) of the leak detector test. |
WavelengthVerTest |
Performs a wavelength verification via the Holmium Oxide Filter. |
WavelengthVerTestDate |
Reports the date of the wavelength verification. |
WavelengthVerTest Result |
Reports the result (pass or fail) of the wavelength verification. |