Full_Scale (Analog Output)
Instrument Type: |
Detector |
Related Commands: |
|
Description: |
The Full_Scale command sets the voltage for a full-scale analog output. For example, if 0.1V is selected and a recorder is connected to the detector’s analog output, the recorder’s full-scale response will be at 0.1V. For the AD20 detector, the full-scale voltage is fixed at 1V. The voltage output of a full-scale detector response to use depends on the recording device to which the analog output is connected. For example, if the analog output is connected to a device that accepts input voltages up to 1 V, select a voltage output of 1000 mV. |