Calibration Examples
Several Standards with Several Substances Each

Calibrations are usually performed using several standards. They often contain more than one substance to be calibrated. The simplest case is two standard samples that contain two substances each. In this case, you can determine the slope and the offset and/or curvature of the calibration curve (depending on the number of acquired data points).

Example:

You want to determine the amount of substances A and B in two samples, Sample I and Sample II. Two standard samples, containing different levels of Standard 1 and Standard 2, are available. Both standard samples contain substance A and substance B. For each sample, an injection volume of 20 µl is injected by an autosampler. The samples are located at the autosampler positions 1 (Standard 1), 2 (Standard 2), 3 (Sample I), and 4 (Sample II).

Sample List

The sequence appears as follows (for more information, refer to Creating and Managing Files and Data  Creating a Sample List (Sequence)):

QNT Method/Peak Table Tab

In Standard 1, for example, the concentration of substance A is 10.2 mg/l and the concentration of substance B is 20.1 mg/l. Standard 2 contains 30.5 mg/l of substance A and 49.7 mg/l of substance B. Thus, you have to create the following peak table in the QNT Editor (For more information about the editor, refer to Data Representation and Reprocessing  The QNT Editor.)

Remove the default Amount column and add two new Amount columns (one for each standard level):

 Note:

The selected option will apply to all sequences that use the current QNT File. For example: You select the Vial Number option, and then assign vial number 5 to Amount column B. From now on, vial number 5 (regardless of its contents) will be assigned to Amount column B in every sequence that uses this QNT File.

 Note:

If one of the substances is not available in the standard sample, the corresponding cell in the Amount column should be left empty.

If you are using several standards for calibration, you may prefer to select a Calibration Type other than Lin (Linear without Offset).

Using these settings, Chromeleon automatically calculates the concentrations of substance A and substance B in the two unknown samples.

For an overview of the calibration options provided by Chromeleon, refer to  Calibrating.