Installing and Controlling Thermo
Scientific Devices
Thermo Scientific TriPlus 100 LS Autosampler
What is required?
Hardware Installation
Module Settings
Prior to installing the autosampler in Chromeleon make sure that the signal cables between autosampler and GC are connected correctly. For connection details, refer to the Pre-Installation Guide and User Guide.
To connect the autosampler with the Chromeleon Server PC
First, install Chromeleon software, and then connect the autosampler to the LAN network. The TriPlus 100 LS autosampler is shipped with a default IP address. For information on the default IP address, refer to the Hardware Manual or check the IP address entered on the (Virtual) Handheld Controller.
Tip:
For better reliability, Thermo Fisher Scientific recommends connecting the autosampler to a separate device LAN and not to the office LAN. Note that an additional network card is required when using a device LAN.
Detailed instructions on how to connect the autosampler to the LAN network are provided in the User Guide and the Hardware Manual provided with the device. General information on connecting modules via LAN is provided in topic .
Software Installation
In the Chromeleon Server Configuration, add the Driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the Configuration dialog box for the autosampler opens automatically.
On each Configuration tab page, select the desired settings. For information about the settings on a page, click the Help button or press F1.
Firmware Update
If the version of firmware installed in the autosampler is not compatible with Chromeleon, for example, after a firmware upgrade on the module or a replacement with a newer model, the configuration is mismatched. When connecting a driver to a module that is not compatible, the connection is aborted due to a firmware version mismatch.
Caution:
Do not downgrade the firmware! Downgrading the firmware to a previous version may result in loss of functionality or malfunctioning of the device.
The autosampler is shipped with the most recent version of firmware at the time. When configuring an existing TriPlus 100 LS into a Chromeleon timebase, if the installed firmware version is below the minimum version supported by Chromeleon, Chromeleon will not be able to connect to the autosampler and an error is logged to the audit trail. Thermo Fisher Scientific recommends to upgrade to the latest firmware version. In this case, contact your Thermo Scientific Service Representative.
Note:
Before performing a firmware update, check the firmware version installed in the autosampler. The firmware version is displayed in the Commands view in the Chromeleon Client, at the Expert display filter level. For details, see The PGM Editor The Commands View.
Further Information
For detailed installation instructions, refer to the manuals and/or guides that are shipped with the autosampler.
If you want to operate the autosampler with two GCs, refer to
Thermo Scientific TriPlus RSH / 100 LS Autosampler: Clone Mode.
For an overview of the different Thermo Scientific modules for which drivers are available in Chromeleon, refer to