Installing and Controlling Third-Party
Devices
Jasco 980 and 1580 Pumps: Installation
Device Type: |
HPLC pump |
Device Driver: |
Jasco: 980/1580 Pump |
Supported Hardware Options: |
Both as HPG and LPG |
What is required?
License: |
Timebase Class 3 |
Connection: |
|
Hardware |
A free RS-232 port. Tip: Firmware version 2.60 or higher is required to control the Jasco 980 pump. Firmware version 2.15 or higher is required to control the Jasco 1580 pump. |
Jasco980.pan |
Hardware Installation
Software Installation
Install the Device Driver in the Chromeleon Server Configuration. (For more information about the program, refer to Software Installation and Communication The Server Configuration Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the pump to the timebase. Select Add Device on the Edit or context menu, and then select Jasco from the Manufacturers list box and 980/1580 Pump from the Devices list box.
The following settings are required:
On the General tab page, select the COM port to which the pump is connected. In addition, disable the Demo Mode.
On the Head Type & Limits tab page, specify the pump type and determine the allowed maximum pressure.
On the Solvents tab page, specify the number and name(s) of the solvents delivered by the pump.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the Jasco 980 or 1580 pump, connect the Jasco980.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
The Jasco 980 pump is suitable for Low-Pressure Gradients (LPG) and High-Pressure Gradients (HPG):
LPG: Chromeleon controls the pump. The pump controls the solvent mixer, which is connected to the pump's Low Pressure Grad output.
HPG: Chromeleon controls pump no. 1. This pump controls additional pumps via LC-Net connections.
Notes:
Due to the slow communication between Chromeleon and the pump, the execution of the flow command is delayed for several seconds. However, a suitable program can compensate this.
If the pressure reading is below the lower pressure limit, the pump will not be turned off for approximately two minutes because the tolerance time of the pump is correspondingly long.
Further Information
Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different Jasco instruments that can be controlled by Chromeleon, refer to Jasco.