Installing and Controlling Third-Party
Devices
Agilent HP 1050 HPLC System: Installation
Device Type: |
HPLC System |
||
Device Drivers: |
Gradient Pump: Agilent or HP: 1050 Gradient Pump (supports four eluents and two relays) Autosampler: Agilent or HP: 1050 Autosampler (supports two relays) UV detector: Agilent or HP: 1050 UV VWD Detector |
||
Hardware
Options |
Device |
Agilent Part No. |
Validated with firmware version |
Gradient pump |
HP 79852 |
3.2 |
|
|
Autosampler |
HP 79855 |
4.2 |
|
Single-channel |
HP 79853A or |
3.02 or |
What is required?
Hardware Installation
Software Installation
Install the Device Drivers for the Agilent (or HP) 1050 HPLC System in the Chromeleon Server Configuration. (For more information about the program, refer to Software Installation and Communication The Server Configuration Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the HP 1050 Gradient pump to the timebase. Select Add Device on the Edit or context menu, and then select Agilent (or HP) from the Manufacturers list box and 1050 Gradient Pump from the Devices list box.
The following settings are required:
On the General tab page, select a free COM port.
On the Solvents tab page, specify the number and names of the eluents that are delivered by the pump.
On the Relays tab page, check a box to enable or disable the corresponding relay.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
Follow step 3 above to add the Agilent (or HP) 1050 Autosampler to the same timebase.
The following settings are required:
On the General tab page, select a free COM port.
On the Rack tab page, select the check box of the used rack.
On the Relays tab page, check a box to enable or disable the corresponding relay.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
In addition, add the Agilent (or HP) 1050 UV VWD Detector to the timebase.
The following settings are required:
On the General tab page, select a free COM port.
On the Signals tab page, accept the enabled channel unless you do not want to record a signal.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the HP 1050 HPLC system connect the HP1050S.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Nothing particular known.
The User Help section provides information about the different commands that the UV detector of the 1050 HPLC System supports; refer to Commands and Tips for Third-Party Devices Agilent 1050 HPLC System: UV Detector.
Troubleshooting
If errors occur during operation, check whether a START/STOP event cable is installed. Remove the cable.
Also, refer to Commands and Tips for Third-Party Devices GPIB-Connected Devices: Troubleshooting in the User Help section.
Further Information
Note:
The Thermo Scientific device drivers for
HP 1050 HPLC System have been tested successfully with the following firmware
versions:
HP 1050 GP rev 3.2 and HP 1050 S rev 4.2.
If you have problems with firmware versions lower than the ones indicated, a firmware update might help.
Tip:
For more installation details, refer to the manual for the instrument.