Tab Dialog Box:
Options
Enter options for the AS-DV autosampler.
Option |
Description |
|
|
Valve |
|
Injection |
This option is selected automatically if an optional 6-port or 10-port high-pressure valve is installed in the AS-DV. When selected, the valve functions as the sample injection valve. |
Auxiliary |
Select this option if an optional 6-port or 10-port high-pressure valve is installed and you want to use it for a function other than sample injection (for example, for bleeding sample to waste before an injection or for loading a concentrator column). |
None |
This option is selected automatically if an optional valve is not installed. |
Label |
Specifies the name used to identify the optional valve in autosampler commands. InjectionValve and AuxValve are the default labels. You can enter a different name if desired. |
Default Injection Parameters |
Chromeleon uses the following formula to calculate the total volume of sample delivered to the injection valve: (Injection Volume + Delay Volume) * Flush Factor The injection volume (typically, the loop volume) is entered in the Sequence. |
Delay Volume |
The volume from the tip of the sampling needle to the injection valve. The default is 125 mL. |
Flush Factor |
Determines how much excess sample will be flushed through the loop; expressed as a multiple of the total of the injection volume and the delay volume. For example, if the injection volume entered in the sequence is Notes:
|
|
|
Rinse Vial Range |
Specify a range of carousel vial positions that will be used to hold rinse vials. When performing a rinse, the AS-DV takes the rinse liquid from the vial in the Start Position. If the start position vial is empty, the next vial in the range is used, and so on. Start Position: The first rinse vial position. End Position: The last rinse vial position. Note: To use another vial position for the rinse vial, enter the position in the Program. |
|
|
Sample Overlap |
Select this check box to enable overlap of samples. When Sample Overlap is enabled, the AS-DV starts the next sample in a sequence while data acquisition is occurring for the current sample. The AS-DV commands that occur before the EndSamplePrep command in the program are overlapped. Note: For sample overlap to occur, the BeginOverlap command must also be present in the Program. Chromeleon automatically adds the BeginOverlap and EndSamplePrep commands when you use the Program Wizard to create a program (see The Program Wizard). The User Help section provides more information; refer to Practical Tips for Device Control: Autosampler Control Overlapping Samples. |