Baseline Point

 

Location:

Detection tab page
Report Variable (Detection Parameters)

Type:

Command

Dimension:

---

Value Range:

---

Default:

---

Related Parameters:

Lock Baseline

Valley to Valley

Inhibit Integration

Type (Peak Type)

Description:

The Baseline Point detection parameter defines a baseline point at the indicated time.

 Tips:

When setting a baseline point, keep in mind that this point will be valid for all chromatograms that are evaluated with the respective QNT Method. If in one of these chromatograms a peak maximum occurs by coincidence at the time of your hard entered baseline point, the peak maximum will be defined as base point and the peak will not be detected.

This command has priority over Lock Baseline and Valley to Valley, but not over Type (Peak Type) from the peak table.

 

For information about how to use individual detection parameters, refer to  Defining Detection Parameters.