Valley to Valley
Location: |
Detection tab page |
Type: |
Switch |
Dimension: |
--- |
Value Range: |
On; Off |
Default: |
Off |
Description: |
Usually, an algorithm determines the baseline automatically. The Valley to Valley parameter enables baseline treatment from valley-to-valley, that is, from peak minimum to peak minimum, in a series of non-resolved peaks. |
Function: |
When the parameter is enabled, the baseline below non-resolved peaks leads from peak end to peak end. |
Notes: |
The end of the last non-resolved peak in a series of peaks is always a baseline point since the chromatogram itself forms the baseline outside the peak. |
|
The Valley to Valley parameter is invalidated by Lock Baseline. |
|
The Type (Peak Type) criterion of the peak table has absolute priority. |
For information about how to use individual detection parameters, refer to Integrating Chromatograms and Identifying Peaks Defining Detection Parameters.