Mix
Instrument Type: |
|
Related Commands: |
|
Description: |
The Mix command causes the autosampler to mix defined injection volumes (see Volume) from several sample vials (see Position) in a different sample vial. |
Function: |
When the action is completed, the autosampler returns the Sampler.Ready signal. Example (ASI-100 Series): |
|
0.000 PrepVial BA1 0.000 PrepVolume 10 0.000 Sampler.PrepVial Position + GA1 - RA1 0.000 PrepSubject PrepVial 0.000 Mix
|
|
The above command sequence mixes 10 µl from a vial in the green segment with 10 µl of a vial in the red segment in the corresponding vial of the blue segment. Example (AS/AS50): |
|
Sampler.Pipet Volume = 20.0, SourceVial = 1, DestinationVial = 10 Sampler.Mix SourceVial = 10, NumberOfTimes = 5, Volume = 30.0 |
|
In the above command sequence, 20 µl is pipetted from vial 1 and delivered to vial 10. The contents of vial 10 is mixed by drawing in and expelling 30 µl of the vial contents. The mixing cycle is repeated 5 times. You can also use the PrepSpeed (ASI-100) or SyringeSpeed (AS/AS50) command to determine the draw speed of the syringe. |
Tip:
|
When using highly viscous liquids or liquids with a low boiling temperature, allow the autosampler more time to draw the exact volume exactly without bubbles. |
For more information, refer to Practical Tips for Device Control: