IC Control
Controlling a DX-120 Ion Chromatograph
The following commands are available for controlling DX-120 operating functions:
Controlled AC |
Turn the AC power outlet on the DX-120 rear panel on and off. This provides on off control of an external accessory connected to the outlet. See the DX-120 operator’s manual for cabling instructions. |
Column/Eluent |
If the DX-120 is in Column Mode, the Column command sets the flow path to column A or column B. If the DX-120 is in Eluent Mode, the Eluent command selects the eluent reservoir (A or B). Refer to Selecting the DX-120 Mode. Note: These commands are active only if the DX-120 is equipped with the dual-column option. |
Data Collection Rate |
Sets the Data Collection Rate of the detector. |
Eluent Pressure |
Turns the pressure to the eluent reservoir(s) on and off. |
Pressure Unit |
Selects the units of pressure to use (psi or MPa). |
Pump |
Turns the pump on and off. |
SRS |
Turns the current supplied to the Self-Regenerating Suppressor (SRS) on and off. |
These commands can be included in a Program, executed directly from the Commands dialog box (to do so, select Command… on the Control menu), or linked to a control on the control panel.
To link a command to a control, place an edit box or switch control on the control panel (see Layout Toolbar). Then, link the control to an object property. For example, to turn the DX-120 pump on and off from the control panel, link the Pump object property to a switch control. For more information, refer to How to: Controlling Devices from the Control Panel Linking a Control to a Device.
The corresponding commands in the program are as follows:
0.000 |
ControlledAC |
=On/Off |
0.000 |
Column |
=A/B |
; |
Only for a DX-120 equipped with the dual-column option. |
0.000 |
Data_Collection_Rate |
=Value [Hz] |
0.000 |
Eluent |
=A/B |
; |
Only for a DX-120 equipped with the dual-column option. |
0.000 |
EluentPressure |
=On/Off |
0.000 |
PressureUnit |
=psi/MPa |
0.000 |
Pump |
=On/Off |
0.000 |
SRS |
=On/Off |