GC Control
Injecting Two GC Samples Simultaneously (6890)

To double the sample throughput, the 6890 GC supports the dual-operation mode, which lets you simultaneously inject two samples from different positions. The following conditions must be met:

After installing the GC, follow the steps below to set up the dual-operation mode.

Creating two PGM Files

Create a PGM File for each timebase, observing the following:

Creating a sequence and a batch for each timebase

It is not possible to simultaneously inject two samples from the same vial (except if UseTray = No).

The batches do not have to include the same number of samples. If all samples of one batch are processed, sample processing will continue for the remaining samples of the other batch.

 Tip: 

You can add new samples or sequences to the finished batch and restart the batch. Sample processing is resumed when the next sample of the running batch is started.

 

Starting batches

In dual operation mode, a batch must be started on both timebases simultaneously:

Examples

a) To analyze a pair of sequences (e.g., Front1 and Back1) simultaneously on two timebases (e.g., Front and Back) and then run another sequence (Front2) only on the Front timebase:

b) Use the following settings if you want to process samples in dual operation mode whenever possible but also analyze them one after the other if required:

c) If you want to run a batch only on one timebase (e.g., on the Front timebase), select one of the following alternatives:

--Or--

In both cases, the batch is started immediately and does not wait for a batch on the Back timebase.

For information about how to install the 6890, refer to  Agilent 6890 GC: Installation.