Installing and Controlling Third-Party Devices
Agilent 6890/7890 GC: Dual Operation

The 6890 and 7890 GCs support dual operation. This lets you simultaneously inject two samples from different sample positions, thus doubling the sample throughput. Make sure that the following conditions are met:

What is required?

License:

Timebase Class 1 License
(Although the GC is installed in two timebases, as a shared device, only one license is required.)

Devices:

6890 GC

7890 (7890A, 7890A+ and 7890B GCs)

Columns:

Two

Injectors:

6890 GC: Front and back injector

7890 GC: Front and back injector or two gas sampling valves

Inlets:

Front and back inlet

Detectors:

Front and back detector

 Note: 

Usually, two identical columns, injectors, inlets, and detectors will be used.

 

Installation in the Server Configuration

For information about the program, refer to Software Installation and Communication  The Server Configuration Program.

To install and configure the 6890 GC for dual operation, follow the steps below:

  1. Create two timebases, e.g., 6890_A and 6890_B.

  2. Install the GC in one timebase, e.g., 6890_A.

  3. On the Options tab page, specify the timebase settings:

     

To install and configure the 7890 GC for dual operation, follow the steps below:

  1. Create two timebases, e.g., 7890_A and 7890_B.

  2. Install the GC in one timebase, e.g., 7890_A.

  3. On the Timebase tab page, specify the timebase settings:

    Note: Use the Valve tab page to assign gas sampling valves to the appropriate Front or Back Inlet. The main device, autosampler, injectors, inlets, oven, and multi position valve (if a gas sampling valve is installed) are assigned automatically to the correct timebase.

 Tip:  

Assign identical names to the Front and Back devices if you are using the same device type. In this case, you can use the same program on both timebases and need to adapt only the column name. For example, you can use the following names:

Front Inlet = Back Inlet = Inlet

Front Injector = Back Injector = Injector

Front Detector = Back Detector = Detector

 

For more information about dual operation on the 6890 GC, refer to:

 Commands and Properties for Dual Operation (6890) and  Injecting Two GC Samples Simultaneously (6890).

 

For more information about dual operation on the 7890 GC, refer to:

 Injecting Two GC Samples Simultaneously (7890).