Commands and Tips for Thermo Scientific Devices
Thermo Scientific Surveyor Plus Series: Pump

The following commands and properties are available for the Surveyor Plus pump (please note that the display Filter level determines which commands and properties are displayed).

Also, refer to the information in the manual shipped with the instrument.

 

General Commands and Properties:

Command/Property

Description

Connect

Connects the instrument to Chromeleon.

Connected

Indicates whether the instrument is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect).

DegasserStatus

Indicates the status of the degasser.

Disconnect

Disconnects the instrument from Chromeleon. Also see Connect.

DriverID

Unambiguous driver name (read-only).

DriverVersion

Driver version (read-only).

FirmwareVersion

Indicates the firmware version of the instrument (read-only).

FirmwareVersionConverter

Indicates the firmware version of the converter board (read-only).

DriverManufacturer

Manufacturer of the driver (read-only).

ModelNo

Reports the instrument's model number (read-only).

ModuleManufacturer

Manufacturer of the hardware module (read-only).

PumpStatus

Indicates the status of the pump.

Ready

Indicates that the instrument is ready for injection.

RunTime

Displays the elapsed method run time.

SerialNo

Indicates the serial number of the instrument (read-only).

 

Flow and Pressure Commands and Properties:

Command/Property

Description

%A, %B, %C, %D

Percentage of each solvent component.

*.Equate

Name of the component (A, B, C, or D).

*.Value

Current value [%] for components A, B, C, or D (read-only).

ClearError

Clears the error condition of the pump.

Compressibility

Specifies the compressibility of the eluent.

Flow

Sets the flow rate [ml/min].

*.Nominal

Reports the nominal flow.

*.Value

Reports the current flow (read-only).

Pressure

Indicates the pump pressure.

*.LowerLimit

Sets the lower pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range.

*.UpperLimit

Sets the upper pressure limit. The system stops the sample batch and starts emergency handling if the value is outside the set range.

*.Value

Current pump pressure (read-only).

PumpOn

PumpOff

Turns the pump on or off.

 

Purge Commands and Properties:

Command/Property

Description

Purge

Starts the purge operation.

PurgeStop

Stops the purge operation.

PurgeTime

Sets the duration of the purge operation.

PurgeValve

Indicates the valve position during the purge operation.

 

Commands and Properties Related to Wear Part and Seal Replacement:

Command/Property

Description

LeftSealVolume

Tracks the volume pumped past the left seal.

MoveToSealReplacement

Moves the piston to a maintenance position that allows replacement of the seal. The command can only be executed if no method is running. After execution of the command, the pump and any running program is stopped.

OperationTime

Tracks the operating time of the pump.

ResetLeftSealVolume

Resets the counter for the LeftSealVolume property to 0. Execute this command after replacement of a piston seal.

ResetRightSealVolume

Resets the counter for the RightSealVolume property to 0. Execute this command after replacement of a piston seal.

ResetOperationTime

Resets the counter for the OperationTime property to 0.

Execute this command after replacement of the motor.

RightSealVolume

Tracks the volume pumped past the right seal.

 

Troubleshooting

The injection is stopped with an Abort message.

The problem occurs when the runtime of the program is set to 1.23 minutes. The reason is an error in the pump's firmware version 3.02.

Solution: Change the run time of the program, for example to 1.22 or 1.24 minutes.

 

For an example for an emergency program, refer to  Thermo Scientific Surveyor Plus Pump: Emergency Program

For information about how to install the instrument, refer to Hardware Installation  Thermo Scientific Surveyor Plus LC Pump: Installation.