Commands and Tips for Third-Party Devices
Shimadzu LC-10A and LC-2010 HPLC Systems: System Settings

The table below lists the commands and properties that are available for the Shimadzu LC-10A and LC-2010 HPLC Systems (please note that the display Filter level determines which commands and properties are displayed).

 Note: 

The names of the Shimadzu commands and properties may be slightly different, depending on the version of your system.

 

Chromeleon

Shimadzu

Values

Description

Activated

 

No, Yes

Activates or deactivates the instrument or system. Reactivate the system if it was deactivated, e.g., because the pressure was below the pressure limit.

 Note:

Activating the system automatically enables temperature control and turns the pump(s) on (Motor = On).

ExternalStart

EXT_S

None,
AllRuns,
InjOnly

Determines which program starts are signalled by the START output (relay 1 on the system controller).

Status

STATUS

 
 

PURGE

PAUSE

HOLD

RUN

ERROR

WAIT_OVEN

OVEN_NOT_READY

TRANSMIT

WAIT_DP

PRETREAT

STOP

RINSE

READY

Indicates the overall instrument or system status. The instrument or system is:

purging.

in Pause mode.

in Hold mode.

running.

An error has occurred.

waiting for the column oven.

The oven is not yet ready for operation.

transmitting data.

waiting for INJ_IN.

busy with sample preparation

stopped.

rinsing.

ready for operation.

WaitForRemoteStart

INJ_IN

Run, Sync

Run: The autoinjector injects the samples immediately after drawing them.

Sync: The injector waits for the manual injection signal before injecting the sample (read-only).

 

For information about how to install the Shimadzu HPLC Systems, refer to Hardware Installation  Shimadzu LC-2010 HPLC System: Installation and  Shimadzu LC-10A/LC-20A HPLC System: Installation in the Administrator Help section.