Commands and Tips for Third-Party Devices
Shimadzu LC-10A and LC-2010 HPLC Systems: Column Oven
In addition to the standard oven commands (see Commands for Controlling Dionex Devices Dionex Flow Manager and Thermostatted Column Compartments), the commands and properties from the table below are available (please note that the display Filter level determines which commands and properties are displayed):
Note:
The names of the Shimadzu commands and properties may be slightly different, depending on the version of your column oven.
Chromeleon |
Shimadzu |
Values |
Description |
AmbientTemperature |
ROOM |
5-85°C |
(Only for the vp module) Indicates the ambient temperature (read-only). |
TempCtrl |
OVEN |
On/Off |
Enables or disables temperature control. Turning off temperature control is not allowed in sample programs. |
Temperature.Nominal |
OVEN.T |
CTO-10A: 4-80°C LC-2010: 4-60°C |
Temperature setpoint for the column oven |
Temperature.Upper |
T.MAX |
CTO-10A: 5-85°C LC-2010: 5-65°C |
Sets the upper temperature limit. Chromeleon aborts the sample batch and starts emergency handling if the current value exceeds the temperature limit. |
Temperature.Value |
TEMPm |
CTO-10A: 4-85°C LC-2010: 4-65°C |
Current temperature of the column oven |
Tips:
You cannot set Temperature.Nominal and TempCtrl for retention times <0.000 min.
Each program may include only one Temperature.UpperLimit setting. In addition, the setting must be entered at the time t = 0.000 min. Otherwise, an error message appears during the Ready Check.
Setting Temperature.Nominal automatically sets TempCtrl to On. If TempCtrl was set to Off before, Log.TempCtrl=On appears in the Audit Trail.
Upon each Connect command, the value for Temperature.Nominal is automatically set to 40 °C, regardless of whether a user-defined value has been set.
For information about how to install the Shimadzu HPLC Systems, refer to: