Commands and Tips for Third-Party Devices
Perkin Elmer Autosystem (XL) and Clarus GCs: Detectors
The following commands and properties are available for detector control (please note that the display Filter level determines which commands and properties are displayed):
Property |
Range/Values |
Description |
Attenuation |
1, 2, 4, 8, 16, 32, 64, (and 128, 256, … 65536 if Output is set to Recorder in the Server Configuration program)) |
Sets the detector attenuation.² |
AZCompensation |
On or Off |
Enables or disables the Autozero option.² |
DetectorStatus |
Clarus 400/500 and Clarus 480/580: NotInstalled, Ok, FlameIgnition, FlameOut, BackgroundComp, FilamentOff, Unknown Clarus 600/680: NotInstalled, Ok, FlameIgnition, FlameOut, FlameCheck, BackgroundComp, FilamentOff, IgnitionFail, Unknown |
Reports additional information about the status of the detector, especially about the status of the flame. If the flame is out, an error message is displayed. Note: To monitor the temperature of the detector, please refer to property Status below. |
DetectorType |
Output, FID, ECD, |
Indicates the detector type (read-only). Select the detector type in the Server Configuration program: FID: Flame Ionization Detector |
FilterTimeConstant |
50 ms (not for the ECD), 200ms, 800ms |
Specifies the time constant for the noise filter of the related detector. |
FlameThreshold |
0.00-9.99 mV |
Flame threshold for evaluating whether the detector flame is lit. If the signal exceeds this value, the GC assumes that the flame is lit. Otherwise, the flame is ignited automatically. |
MaxAutoStep |
0.1 to 5.1 s |
Specifies the maximum step for the auto step mode. The default setting is 5.1 s. |
Range |
1, 20 (FID, NPD) 1 (ECD) -4 - 4 (TCD) |
Sets the detector output range.² |
Ready |
Ready or NotReady |
Indicates whether the detector is ready (read-only; level: Advanced or higher). If a PPC is installed, the status of the gas flows is taken into account, also. |
Retention |
0 - 10^8 min |
Indicates the retention time. |
Signal |
-10000 - 10000mV |
Indicates the current signal value (read-only). |
Status |
None, Ready, NotReady, FilamentOver, Unknown |
Indicates the detector status (read-only). |
Step |
0.01 to 4.80 s |
Sets the step. The default setting is Auto (recommended). With this setting, the optimum step is selected dynamically. |
TempCtrl |
On/Off |
Turns the heating component of the detector on or off. |
Temperature |
100 to 450°C 100 to 350°C (TCD) |
Properties related to the detector temperature. Note: If a temperature is set, TempCtrl is automatically set to On. |
Temperature. |
Sets the nominal temperature. |
|
Temperature. |
Indicates the current temperature (read-only). |
² If the value is entered manually or at time t=0, the value used at the beginning of the run is adapted.
Tip:
If a value is entered at another time, this value remains valid until the next setting or the beginning of the next GC run.
In addition, Chromeleon supports the following commands and/or properties if the Autosystem GC includes a PPC:
Property |
Values |
Description |
<Gasname> |
0–500 ml/min |
<Gasname> is the gas specified on the Detectors tab page in the Server Configuration program, e.g., He, N2, H2, Ar/CH4, or Air. Note: You can enter the gas name on the Detectors tab page in the Server Configuration program. |
<Gasname>. |
Sets the nominal flow for the associated gas. |
|
<Gasname>. |
Indicates the current flow for the associated gas (read-only; level: Advanced or higher). |
|
<GasName>_Ctrl |
On/Off |
|
<GasName>_GasType |
He, N2, H2, Ar/CH4, Air |
Indicates the gas type in the PPC module (read-only; level: Advanced or higher). Note: The gas type is set in the Server Configuration program. |
<GasName>_Status |
NotUsed, Ready, NotReady, Alarm |
Indicates the current status in the PPC module (read-only; level: Advanced or higher). |
For installation details, refer to: