Commands and Tips for Third-Party Devices
Merck Hitachi Autosamplers
Supported Autosamplers
For Autosampler… |
…controlled via… |
…use device driver |
AS2000 |
RS-232 |
Merck Hitachi AS2000 Autosampler |
AS4000 |
RS-232 |
Merck Hitachi AS4000 Autosampler |
L7250 |
RS-232 |
Merck Hitachi L7250 Autosampler |
L7250 |
D-Line |
Merck Hitachi L7200/L7250 AS (D-Line) |
L7200 |
D-Line |
Merck Hitachi L7200/L7250 AS (D-Line) |
Supported Device Properties
In addition to the standard autosampler commands (see Dionex Autosamplers), the Merck Hitachi autosampler support the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
Property |
AS2000 |
AS4000 |
L7250 |
L7200 |
InjectPortWashs |
0-20 |
0-20 |
0-20 |
0-20³ |
InjectPortWashSpeed |
1 (Low) ... |
1 (Low) ... |
1 (Low) ... |
1 (Low) ... |
NeedleWashs |
1-20 |
1-20 |
1-20 |
1-20³ |
NeedleWashSpeed |
1 (Low) ... |
1 (Low) ... |
1 (Low) ... |
1 (Low) ... |
WasteVolume |
Not supported |
Not supported |
0-4000.0 µl |
0-4000.0 µl² |
WashVolume |
10.0-100.0 µl |
10.0-1000.0 µl |
10.0-1000.0 µl |
|
LeadVolume |
0-100.0 µl |
0-1000.0 µl |
0-1000.0 µl |
0-500.0 µl² |
RearVolume |
0-100.0 µl |
0-1000.0 µl |
0-1000.0 µl |
0-500.0 µl² |
FeedVolume |
0-100.0 µl |
0-1000.0 µl |
0-1000.0 µl |
0-500.0 µl² |
DeadVolume |
0-100.0 µl |
0-1000.0 µl |
0-1000.0 µl |
|
Syringe |
0-100.0 µl |
0-1000.0 µl |
0-1000.0 µl |
|
InjectionMethod |
0 (Compatibility) |
0 (Compatibil.) |
0 (Compatibil.) |
1 (Cut) ... |
SyringeSpeed |
1 (Low) ... |
1 (Low) ... |
1 (Low) ... |
|
NeedleDownSpeed |
1 (Slow) ... |
1 (Slow) ... |
1 (Slow) ... |
1 (Slow) ... |
VolSyringe |
1 (100µl) - |
1 (100 µl) -5 (5000 µl) |
1 (100 µl) -5 (5000 µl) |
Not supported |
PumpPlungerWash |
Not supported |
Not supported |
0 (No) or |
0 (No) or |
VolOptionalSyringe |
Not supported |
1 (100 µl) - 5 (5000 µl) |
Not supported |
Not supported |
NumberOfDilutors |
Not supported |
1 or 2 |
Not supported |
Not supported |
Dilutor |
Not supported |
1 or 2 |
Not supported |
Not supported |
RackCode |
1 - 16 |
1 - 16 |
1 - 16 |
Not supported |
1 - 500.0 µl |
1 - 5000.0 µl |
1 - 5000.0 µl |
1 - 500.0 µl |
|
1 - 100 |
1 - 200 |
1 - 200 |
1 - 200 |
² If one of these properties is set, all other settings that are marked with ² are set, too. Check the parameters sent to the autosampler with extra care.
³ If one of these properties is set, all other settings that are marked with ³ are set, too. Check the parameters sent to the autosampler with extra care.
Supported Device Commands
Command |
AS2000 |
AS4000 |
L7250 |
L7200 |
Supported |
Supported |
Supported |
Supported |
|
Disconnect |
Supported |
Supported |
Supported |
Not supported |
Aspirate (see Draw) |
0 - 500.0 µl, Speed |
Dilutor, 0 - 5000.0 µl, Speed |
0 - 5000.0 µl, Speed |
Not supported |
0 - 500.0 µl, Speed |
Dilutor, 0-5000.0 µl, Speed |
0-5000.0 µl, Speed |
Not supported |
|
GotoTube |
1 - 100, RackCode |
1 - 200, RackCode |
1 - 200, RackCode |
Not supported |
GotoWashPort |
Supported |
Supported |
Supported |
Not supported |
GotoInjectPort |
Supported |
Supported |
Supported |
Not supported |
MoveNeedle |
X:0-163.0 mm Y:0-68.0 mm Z:0-38.0 mm |
X:0-244.0 mm Y:0-163.0 mm Z:0-57.0 mm |
X:0-156.0 mm Y:0-156.0 mm Z:0-57.0 mm |
Not supported |
Height |
Z:0-38.0 mm |
Z:0-57.0 mm |
Z:0-57.0 mm |
Not supported |
GotoSyringePos |
0 - 500.0 µl, Speed |
Dilutor, 0-5000.0 µl, Speed |
0 - 5000.0 µl, Speed |
Not supported |
GotoSValvePos |
0 (Bottle) or |
Dilutor, 0 (Bottle) or 1 (Needle) |
0 (Bottle) or |
Not supported |
GotoValvePos |
0 (Inject) or |
0 (Inject) or |
0 (Inject) or |
Not supported |
Home |
Supported |
Supported |
Supported |
Not supported |
Reset |
Supported |
Supported |
Supported |
Supported |
Wash |
Supported |
Supported |
Supported |
Not supported |
WashPumpPlunger |
Not supported |
Not supported |
Supported* |
Not supported |
Inject |
Supported - |
Supported - |
Supported - |
Supported |
Wash_X |
Supported - |
Not supported |
Supported - |
Not supported |
WashDil_X |
Not supported |
Supported - |
Not supported |
Not supported |
GotoOpValvePos |
Not supported |
Not supported |
0 (Inject) or |
Not supported |
Inject Programs
AS2000 AS4000 L7250 |
Depending on the current value of the Syringe device property, three different programs are internally executed during processing an Inject command. For AS2000 and L7250 the dilutor property always equals 1. |
|
Syringe>0 |
GotoSyringePos SyringePos=0 Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoSValvePos SValvePos=Bottle Dilutor=<current value of property Dilutor> GotoSyringePos SyringePos=<current value of property Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoTube TubeNumber=<current value of property RackCode=<current value of property Aspirate AspirateVol=<current value of property Volume> Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoInjectPort GotoValvePos ValvePos=Load GotoSyringePos SyringePos=0 Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoValvePos ValvePos=Inject |
|
Syringe=0 |
GotoTube TubeNumber=<current value of property Position> RackCode=<current value of property Aspirate AspirateVol=<current value of property Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoInjectPort GotoValvePos ValvePos=Load GotoSyringePos SyringePos=0 Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoValvePos ValvePos=Inject |
|
Syringe<0 |
GotoValvePos ValvePos=Inject |
|
L7200 |
Based on the current parameter settings, the autosampler performs an injection. The Inject command transmits the vial number and the sample volume only. |
Wash Programs
AS2000 AS4000 L7250 |
This is the program that is run when the extended wash command (Wash_X resp. WashDil_X) is executed. For AS2000 and L7250 the dilutor property always equals 1. GotoWashPort GotoSValvePos SValvePos=Bottle Dilutor=<current value of property Dilutor> GotoSyringePos SyringePos=<current value of property Syringe> Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoSValvePos SValvePos=Needle Dilutor=<current value of property Dilutor> GotoSyringePos SyringePos=0 Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoInjectPort GotoSValvePos SvalvePos=Bottle Dilutor=<current value of property Dilutor> GotoSyringePos SyringePos=<current value of property Syringe> Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> GotoSValvePos SValvePos=Needle Dilutor=<current value of property Dilutor> GotoSyringePos SyringePos=0 Speed=<current value of property Speed> Dilutor=<current value of property Dilutor> |
L7200 |
not supported |
More information about the different autosampler types is provided in the following topics:
Merck Hitachi AS2000 and AS4000 Autosamplers: Overview