Commands and Tips for Third-Party Devices
Agilent G1888 Network Headspace Sampler
The following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
General Commands and Properties:
Command/Property |
Description |
Connect |
Connects the device to Chromeleon. |
Connected |
Indicates whether the instrument is connected to Chromeleon, that is, under computer control (also, see Connect, Disconnect). |
Disconnect |
Disconnects the instrument from Chromeleon and activates manual control from the instrument panel. |
DriverID |
Unambiguous driver name (read-only). |
DriverVersion |
Driver version (read-only). |
FirmwareVersion |
Firmware version of the headspace sampler (read-only). |
Manufacturer |
Manufacturer of the driver (read-only). |
ModelNo |
Model number of the headspace sampler (read-only). |
Ready |
Ready indicates that the sampler is ready for operation, i.e., no Autozero started, temperature reached, no errors. NotReady indicates that the sampler is not ready for operation (read-only). |
SerialNo |
Serial number (read-only). |
Special sampler commands
Command/Property |
Description |
CarrierPressure |
Reports the current carrier gas pressure (read-only) |
CarouselPosition |
Moves the carousel to the selected position (1-12). |
*.Nominal |
Sets the position (1-12) the carousel is moved to by the CarouselPosition command. |
*.Value |
Reports the current position and current status of the carousel (read-only): MOVING: Carousel is moving. ERR: Error, e.g., motor encoder senses obstruction or cannot find home position, temperature is out of range. FAIL: Component is not able to work. LOCK: Component is disabled to protect the instrument, e.g., if another component fails. INIT: Initializing or self test state |
CurrentTrayPosition |
Reports the current tray position from the instrument status (read-only). |
ExtractionsPerVial |
Sets the number of extractions per vial in MHE Concentration Mode. |
FirstVial |
Reports the location of the first vial to be analyzed by this method (read-only). |
GCCycleTime |
Sets the GC run time between two injections. |
GCHandshake |
Reports the Handshake parameters of the sampler as defined in the Chromeleon Server Configuration (read-only). |
GCStatus |
Reports the current GC status. |
GlobalReset |
Resets the instrument, including the communications link, to the settings that were valid when the sampler was turned on. |
Inject |
Injects the sample. |
InjectTime |
Sets the time during which the vent valve must be open to allow the sample gas to flow into the GC. |
LastVial |
Reports the location of the last vial to be analyzed by this method (read-only). The number of the last vial must not be lower than the number of the first vial. |
LastVialInjected |
Reports the number of the last vial injected (read-only). |
LoopEquilibrationTime |
Sets the time during which the vent valve must be closed. This allows the analytes in the sample loop to equilibrate to the higher loop temperature and allows the pressure and flow in the loop to stabilize. |
LoopFillTime |
Sets the time required to fill the sample loop with gas. |
LoopTemp |
Sample loop temperature |
*.Nominal |
Sets the nominal sample loop temperature. |
*.Value |
Reports the current sample loop temperature (read-only). |
MHEConcMode |
Select On if you want to use Multi Headspace Extraction (MHE) in Concentration Mode. Select Off to perform multi extraction in single or standard mode. Note: In addition to setting the MHEConcMode setting to "On" or "Off", you must also select the related setting in the "Pos." column of the sample table (sequence) in the Chromeleon Browser to activate a mode: Select different positions in the sample table for all samples to be processed. Standard Mode: MHEConcMode = OFF Select the same position in the sample table for all samples to be processed. Single Mode: MHEConcMode = OFF Select different positions in the sample table for all samples to be processed. |
OvenTemp |
Oven temperature |
*.Nominal |
Sets the nominal oven temperature. |
*.Value |
Reports the current oven temperature (read-only). |
ParIncOvenTempStep |
Increases the oven temperature (OvenTemp) by that many degrees for the second vial and each subsequent vial. |
ParIncVialEquilibrationStep |
Increases the equilibration time (VialEquilibrationTime) by that many minutes for the second vial and each subsequent vial. |
Position |
Sets the vial position. |
PressureUnit |
Sets the pressure unit (kPa, psi or bar). |
PressureValvePosition |
Switches the pressurization valve position. |
*.Nominal |
Allows you to manually set the nominal pressurization valve position (CLOSED or OPEN). |
*.Value |
Reports the current pressurization valve position and status (read-only). |
PressurizationEqTime |
Sets the time required to reach equilibration in the vial. |
PressurizationTime |
Sets the time during which pressurization gas is introduced into the sample vial to build up additional pressure in the vial. |
SampleLifterPosition |
Allows you to manually set the sample lifter position. UP: Raises the lift to move the sample to the injection position. DOWN: Lowers the lifter to move the sample back into the carousel. |
*.Nominal |
Sets the nominal sample lifter position (DOWN or UP). |
*.Value |
Reports the current sample lifter position and status (read-only). |
SampleValvePosition |
Switches the sample valve position. |
*.Nominal |
Allows you to manually set the nominal sample valve position (CLOSED or OPEN). |
*.Value |
Reports the current sample valve position and status (read-only). |
SequenceValvePurgeTime |
Sets the time that the sample valves during the sequence run after each injection to purge the system. |
Shaking |
Sets the agitation level of the oven carousel during the equilibration phase. OFF turns the shaking off. |
ShutterPosition |
Switches the shutter position. |
*.Nominal |
Allows you to manually set the nominal shutter position (CLOSED, MID or OPEN). |
*.Value |
Reports the current shutter position and status (read-only). |
SoftReset |
Resets the instrument to the settings that were valid when the sampler was turned on. However, the current communications link setting is maintained and manual user interaction is not required. |
StandbyValvePurge |
Time the sample and vent valves are switched during standby mode to purge the system. |
*.Nominal |
Sets the time the valves open. |
*.Value |
Reports the current standby valve purge time (read-only). |
TempEquilibrationTime |
The time the sampler waits after the specified temperatures have been reached before it switches into the Ready state. |
TimeRemainingInOven01 -TimeRemainingInOven12 |
Reports the number of minutes until the vial in the specified oven location will be ready for injection (read-only). |
TransferLineTemp |
Temperature of the transfer line to the GC |
*.Nominal |
Sets the nominal transfer line temperature. |
*.Value |
Reports the current transfer line temperature (read-only) |
TrayLifterPosition |
Switches the tray lifter position. |
*.Nominal |
Allows you to manually set the tray lifter movement (UP or DOWN): DOWN: Lowers the lifter to move the sample down into the oven. UP: Raises the lifter to return the sample into the tray. |
*.Value |
Reports the current tray lifter position and status (read-only). |
TrayPosition |
Moves the tray to a selected loading position.. |
*.Nominal |
Allows you to manually set the injection position (1-70) to which the tray is moved when TrayPosition is executed. |
*.Value |
Reports the current tray position (read-only). |
ValveCount |
Counts the number of valve movements (read-only). |
VentValvePosition |
Switches the vent valve position . |
*.Nominal |
Allows you to manually set the nominal vent valve position (CLOSED or OPEN). |
*.Value |
Reports the current vent valve position and status (read-only). |
VialEquilibrationTime |
Sets the time that the vial remains in the oven at the predefined temperature. |
VialInOven01 - VialInOven12 |
Reports the current number of the vial (01-70) in the specified carousel location (01-12) in the oven (read-only). 0 if the location is empty. |
VialPressure |
Reports the current vapor pressure of the vial (read-only). |
VialSize |
Sets the volume [mL] of the vials used. |
For installation details, refer to Installing and Controlling Third-Party Devices Agilent G1888 Network Headspace Sampler: Installation.