Installing and Controlling Third-Party
Devices
Agilent G1888 Network Headspace Sampler: Installation
Device Type: |
Headspace Autosampler |
Device Driver: |
Agilent: G1888 Headspace Sampler |
Supported Hardware Options:
|
No special options available |
Restrictions: |
The following instrument options are not supported:
|
What is required?
License: |
Timebase Class 1 |
Connection:
|
Use an Ethernet LAN cable, which is shipped from Agilent with the autosampler to connect the device to the network. –OR–- Use a standard crossover cable to connect the autosampler directly to the server PC. |
Hardware Prerequisites: |
A free Ethernet LAN port is required. |
Firmware Prerequisites: |
Tested for firmware A.01.08 |
Agilent_HS_G1888 |
Hardware Installation
Device Connection: |
Ethernet LAN connection (it is not possible to connect the autosampler via an RS-232 interface). Tips: For better reliability, Dionex recommends connecting the Headspace autosampler to a separate device LAN and not to the Office LAN. It is not possible to connect the autosampler via an RS-232 interface. Synchronization between the autosampler and the GC can either be realized via a Remote cable connecting the autosampler with the GC or via Chromeleon. Synchronization via remote cable: The Remote cable is provided by Agilent. For information on performing synchronization using a remote cable, refer to 7890 GC: Instrument Synchronization. Synchronization via Chromeleon:
|
Device
Settings |
To control the Headspace autosampler via a LAN connection, make the following settings on the autosampler display:
|
Software Installation
In the Chromeleon Server Configuration, add the driver to a timebase. For details, see Adding, Configuring or Deleting Devices. After you add the driver, the configuration dialog opens automatically.
On each properties tab page, select the desired settings. For information about the settings on a page, click the Help button on a page or press F1.
To control the G1888 Headspace autosampler, connect the Agilent_HS_G1888.pan Control Panel in the Chromeleon client to the corresponding timebase.
Troubleshooting
If you have special questions, please contact Dionex Service.
Further Information
For installation details, refer to the G1888 Network Headspace Sampler Operating Instructions.
For an overview of the different Agilent/HP instruments for which Dionex device drivers are available, refer to Agilent (formerly HP).